246
Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe]:DETector:DELay1
Syntax
[:SENSe]:DETector:DELay1 <numeric>
[:SENSe]:DETector:DELay1?
Description
Sets/Gets the waiting time for analog measurement for 120 Hz Frequency measurement.
Parameters
Query response
{numeric}<newline><^END>
Related commands
[:SENSe]:DETector:DELay2 on page 246
[:SENSe]:DETector:DELay3 on page 247
Equivalent key
sequence
No equivalent key can be used on front panel.
[:SENSe]:DETector:DELay2
Syntax
[:SENSe]:DETector:DELay2 <numeric>
[:SENSe]:DETector:DELay2?
Description
Sets/Gets the waiting time for analog measurement for 1 kHz Frequency measurement.
Parameters
Query response
{numeric}<newline><^END>
Related commands
[:SENSe]:DETector:DELay1 on page 246
[:SENSe]:DETector:DELay3 on page 247
Equivalent key
sequence
No equivalent key can be used on front panel.
<numeric 1>
Description
Waiting time for 120 Hz measurement frequency
Range
0 to 100m
Initial value 1.67m
Resolution
1u
<numeric 1>
Description
Waiting time for 1 kHz measurement frequency
Range
0 to 100m
Initial value 1m
Resolution
1u
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...