![Keysight Technologies E4981A Скачать руководство пользователя страница 223](http://html1.mh-extra.com/html/keysight-technologies/e4981a/e4981a_programming-manual_1961965223.webp)
Chapter 10
221
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:FORMat:STATus:EXTension
Syntax
:FORMat:STATus:EXTension {ON|OFF|1|0}
:FORMat:STATus:EXTension?
Description
This command enables/disables the status extension reporting for Contact check failure.
If the status reported (numeric1) from
is 2 (Detection of
LowC or No Contact), then
can be enabled to find the
source of Contact Check failure. The return value of status determines the source of contact
check failure and are as follows:
• +8: Threshold1 Failure
• +16: Threshold2 Failure
• +128: Signal Level Check failure when LVL COMP = ON and Contact Check = ON.
Parameter
Query response
{1|0}<newline><^END>
Related commands
*TRG on page 169
:FETCh? on page 217
:READ? on page 226
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe] on page 256
Equivalent key
No equivalent key is available on the front panel.
Description
ON or 1
Enables the extended reporting of Contact Check Failure.
OFF or 0
Disables the extended reporting of Contact Check Failure.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...