Keysight Technologies E2655C Скачать руководство пользователя страница 14

14

E2655C Probe Deskew and Performance Verification Kit User’s Guide

 

Calibration for Hand-held Browser Probe Heads

Calibration of the hand-held browser probe heads consists of a vertical calibration 

and a skew calibration. The vertical calibration should be performed before the 

skew calibration. Both calibrations should be performed for best probe 

measurement performance.

Calibration Setup

The calibration procedure requires the following parts.
• BNC (male) to SMA (male) adapter
• Deskew  fixture
• 50 

Ω

 SMA terminator

For the following procedure, refer to 

Figure 7

.

1

Connect BNC (male) to SMA (male) adapter to the deskew fixture on the 

connector closest to the pincher.

2

Connect the 50 

Ω

 SMA terminator to the connector farthest from the pincher.

3

Connect the BNC side of the deskew fixture to the Aux Out of the Infiniium 

oscilloscope.

4

Connect the probe to an oscilloscope channel.

5

Place the positive resistor tip of the browser on the center conductor of the 

deskew fixture between the green line and front end of the pincher. The 

negative resistor tip or ground pin of the browser must be on either of the two 

outside conductors (ground) of the deskew fixture.

6

On the Infiniium oscilloscope in the Setup menu, select the channel connected 

to the probe.

7

In the Channel Setup dialog box select the 

Probe Config...

 button.

8

In the Probe Configuration dialog box select the 

Calibrate Probe...

 button.

9

Select the 

Start Atten/Offset Calibration...

 button and follow the on-screen 

instructions for the vertical calibration procedure.

10

Once the vertical calibration has successfully completed, select the 

Calibrated 

Skew...

 button.

11

Select the 

Start Skew Calibration...

 button and follow the on-screen 

instructions for the skew calibration.

NOTE

Before calibrating the probe, verify that the Infiniium oscilloscope has been calibrated recently 

and that the calibration 

Δ

 temperature is within ±5 °C. If this is not the case, calibrate the 

oscilloscope before calibrating the probe. This information is found in the Infiniium Calibration 

dialog box.

Содержание E2655C

Страница 1: ...Keysight E2655C Probe Deskew and Performance Verification Kit User Guide...

Страница 2: ...ides the Software to U S gov ernment customers under its standard com mercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www key sight com fi...

Страница 3: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 3...

Страница 4: ...4 E2655C Probe Deskew and Performance Verification Kit User s Guide...

Страница 5: ...1131B 1132B 1134B and the InfiniiMode differential probes N2750A N2751A N2752A is done using the E2655C Probe Deskew and Performance Verification Kit The kit contains the following parts SMA male to S...

Страница 6: ...tion for the InfiniiMax 1160B series and 1130B series probes solder in differential probe head and the differential browser probe head However the procedures also apply to all of the different Infinii...

Страница 7: ...nfiniium oscilloscope 4 Connect the probe to an oscilloscope channel 5 To minimize the wear and tear on the probe head the probe head should be placed on a support to relieve the strain on the probe h...

Страница 8: ...uide NOTE To ensure contact pull up on the back side of the pincher to ensure good contact between resistor leads and the deskew fixture Figure 2 Connecting the Probe for Calibration BNC to SMA Connec...

Страница 9: ...front panel 2 Set the volts per division to 100 mV div 3 Set the horizontal scale to 1 00 ns div 4 Set the horizontal position to approximately 3 ns You should see a waveform similar to that in Figur...

Страница 10: ...he Probe Configuration dialog box select the Calibrate Probe button 5 In the Probe Calibration dialog box select the Start Atten Offset Calibration button and follow the on screen instructions for the...

Страница 11: ...ctor farthest from the pincher 3 Connect the BNC male to BNC male cable to the BNC connector on the deskew fixture to one of the unused oscilloscope channels For Infiniium oscilloscopes with bandwidth...

Страница 12: ...e Auto Scale key on the front panel NOTE To ensure contact pull up on the back side of the pincher to ensure good contact between resistor leads and the deskew fixture Figure 5 Connection setup for ve...

Страница 13: ...boxes 16 Select the Start Skew Calibration button and follow the on screen instructions 17 Set the vertical scale for the displayed channels to 100 mV div 18 Set the horizontal range to 1 00 ns div 1...

Страница 14: ...the browser on the center conductor of the deskew fixture between the green line and front end of the pincher The negative resistor tip or ground pin of the browser must be on either of the two outsi...

Страница 15: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 15 Figure 7 Calibration setup BNC to SMA Connector not visible InfiniiMax Probe Pincher Deskew Fixture 50 W SMA Terminator...

Страница 16: ...16 E2655C Probe Deskew and Performance Verification Kit User s Guide...

Страница 17: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 17...

Страница 18: ...This information is subject to change without notice Keysight Technologies 2003 2004 2013 2016 Edition 5 August 2016 E2655 92004 www keysight com...

Отзывы: