background image

Keysight E2307A and E2308A User’s Guide

3

Keysight E2307A Thermocouple Bead Temperature Probe

– Exposed welded thermocouple junction with PTFE insulation wire, 36 inches 

(0.9 m) long, terminated in a 0.75 inch dual-banana plug.

Thermocouple type

: #24 gage ANSI-K type calibration.

Accuracy

: –260 

°

C to –110 

°

C (–436 

°

F to –200 

°

F)   2% of reading

–110 

°

C to 260 

°

C (–200 

°

F to 500 

°

F) 

±

2.2 

°

C (

±

°

F)

Maximum temperature

: 260 

°

C (500 

°

F)

For use with

: Keysight 973A or any other device that accepts type-K 

thermocouples and a dual-banana plug.

Keysight E2308A Thermistor Temperature Probe

– Stainless steel encapsulated thermistor (5000 

Ω

 @ 25 

°

C), 36 inch (0.9 m) 

leads, terminated in a 0.75 inch dual-banana plug.

Accuracy

: –40 

°

C to 0 

°

C (–40 

°

F to 32 

°

F)  

±

0.4 

°

C (0.8 

°

F)

°

C to 70 

°

C (32 

°

F to 158 

°

F) 

±

0.2 

°

C (0.4 

°

F)

70 

°

C to 100 

°

C (158 

°

F to 212 

°

F) 

±

0.3 

°

C (0.6 

°

F)

100 

°

C to 150 

°

C (212 

°

F to 302 

°

F) 

±

°

C (2 

°

F)

Maximum temperature

: 150 

°

C (302 

°

F)

Time constant

: 3 seconds typical in well-stirred water

Isolation

: 42 volts peak max

Stability (for 10 months)

: 25 

°

C <0.01 

°

C

100 

°

C <0.2 

°

C

150 

°

C <1.5 

°

C

Stainless steel probe

: 1.5 in. L x 0.125 in. dia. (38 mm L x 3.2 mm dia.)

For use with

: Keysight 971A/972A/973A/974A and Keysight 3458A. Can also 

be used with the Keysight 34401A with application software.

Содержание E2307A

Страница 1: ...Keysight E2307A Thermocouple Bead Probe Keysight E2308A Thermistor Probe User s Guide...

Страница 2: ...rights customarily provided to the public to use modify reproduce release per form display or disclose commercial computer software or commercial com puter software documentation No additional governm...

Страница 3: ...lug Keysight E2308A Thermistor Temperature Probe Stainless steel encapsulated thermistor 5000 25 C 36 inch 0 9 m leads terminated in a 0 75 inch dual banana plug Accuracy 40 C to 0 C 40 F to 32 F 0 4...

Страница 4: ...th Keysight 970 Series Handheld Multimeters Keysight E2307A thermocouple probe Keysight E2308A thermistor probe Keysight 973A Measurement range 50 C to 260 C 58 F to 500 F Resolution 1 C 1 F Accuracy...

Страница 5: ...C 0 8 F 0 5 C 1 F 1 2 C 2 4 F NOTE Other temperature probes suitable for the Keysight multimeters are available for almost any application Type K thermocouples may be used with the Keysight 973A Some...

Страница 6: ...6 Keysight E2307A and E2308A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 7: ...o change without notice Always refer to the English version at the Keysight website for the latest revision Keysight Technologies 2003 2016 Edition 2 September 23 2016 Printed in Malaysia E2307 90002...

Страница 8: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Keysight E2308A...