Testing Performance
2
1000 X-Series Oscilloscopes Service Guide
33
Test Internal Trigger Sensitivity
Follow these steps to perform the internal trigger sensitivity test:
1
On the oscilloscope, press the
[Default Setup]
key.
2
Press the
[Mode/Coupling]
key; then, press the
Mode
softkey to select
Normal
.
3
Connect the equipment (see
a
Connect the signal generator output to the oscilloscope channel 1 input.
b
Connect a 50 ohm feedthrough termination between the channel 1 input
and the BNC cable.
Table 10
Internal Trigger Sensitivity Specifications
Oscilloscope
Models
Internal Trigger Sensitivity Specification Band wid th
<= 10 MHz
Maximum Band wid th
50 MHz
greater of 0.6 div or 2.5 mV
greater of 0.9 div or 3.8 mV
70 MHz
greater of 0.6 div or 2.5 mV
greater of 0.9 div or 3.8 mV
100 MHz
greater of 0.6 div or 2.5 mV
greater of 1.2 div or 5 mV