Keysight InfiniiVision 2000 X-Series Oscilloscopes 

User's Guide

 

18 Save/Recall (Setups, 

Screens, Data)

Saving Setups, Screen Images, or Data / 229
Recalling Setups, Masks, or Reference Waveforms / 237
Recalling Default Setups / 238
Performing a Secure Erase / 239

Oscilloscope setups, reference waveforms, and mask files can be saved to internal 

oscilloscope memory or to a USB storage device and recalled later. You can also 

recall default or factory default setups.

Oscilloscope screen images can be saved to a USB storage device in BMP or PNG 

formats.

Acquired waveform data can be saved to a USB storage device in 

comma-separated value (CSV), ASCII XY, and binary (BIN) formats.

There is also a command to securely erase all the oscilloscope's non-volatile 

internal memory.

Saving Setups, Screen Images, or Data

1

Press the 

[Save/Recall]

 key.

2

In the Save/Recall Menu, press 

Save

.

3

In the Save Trace and Setup Menu, press 

Format

, then, turn the Entry knob to 

select the type of file you want to save:

Содержание DSO-X 2002A

Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...

Страница 2: ...under the same terms by which the software is customarily provided to the public Accordingly Keysight provides the Software to U S government customers under its standard commercial license which is...

Страница 3: ...ING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyo...

Страница 4: ...els 2 and 4 channel digital storage oscilloscope DSO models 2 8 channel and 4 8 channel mixed signal oscilloscope MSO models Table 1 2000 X Series Model Numbers Bandwidths Bandwidth 70 MHz 100 MHz 200...

Страница 5: ...ial decode packets Digital channels and serial decode cannot be on at the same time The Serial key takes precedence over the Digital key Serial triggers can be used when digital channels are on Math w...

Страница 6: ...s starting on page 119 When setting up triggers or changing how data is acquired see Chapter 10 Triggers starting on page 125 Chapter 11 Trigger Mode Coupling starting on page 147 Chapter 12 Acquisiti...

Страница 7: ...ial Decode starting on page 317 Chapter 25 UART RS232 Triggering and Serial Decode starting on page 335 NOTE Abbreviated instructions for pressing a series of keys and softkeys Instructions for pressi...

Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...

Страница 9: ...sy Viewing 26 Power On the Oscilloscope 27 Connect Probes to the Oscilloscope 28 Maximum input voltage at analog inputs 28 Do not float the oscilloscope chassis 29 Input a Waveform 29 Recall the Defau...

Страница 10: ...ob s coarse fine adjustment setting 57 To position the time reference left center right 58 Searching for Events 58 To set up searches 59 Navigating the Time Base 59 To navigate time 59 To navigate sea...

Страница 11: ...ors 74 Add or Subtract 74 Multiply or Divide 75 Math Transforms 76 FFT Measurement 76 5 Reference Waveforms To save a waveform to a reference waveform location 85 To display a reference waveform 86 To...

Страница 12: ...6 To change the logic threshold for digital channels 96 To reposition a digital channel 97 To display digital channels as a bus 98 Digital channel signal fidelity Probe impedance and grounding 101 Inp...

Страница 13: ...Trigger 127 Pattern Trigger 130 Hex Bus Pattern Trigger 132 Pulse Width Trigger 132 Video Trigger 135 To trigger on a specific line of video 138 To trigger on all sync pulses 140 To trigger on a spec...

Страница 14: ...lloscope Rise Time 159 Oscilloscope Bandwidth Required 160 Memory Depth and Sample Rate 160 Selecting the Acquisition Mode 161 Normal Acquisition Mode 162 Peak Detect Acquisition Mode 162 Averaging Ac...

Страница 15: ...187 Base 188 Overshoot 188 Preshoot 189 Average 190 DC RMS 190 AC RMS 191 Time Measurements 193 Period 193 Frequency 194 Width 195 Width 195 Duty Cycle 195 Rise Time 195 Fall Time 195 Delay 196 Phase...

Страница 16: ...add modulation to the waveform generator output 222 To set up Amplitude Modulation AM 223 To set up Frequency Modulation FM 224 To set up Frequency Shift Keying Modulation FSK 226 To restore waveform...

Страница 17: ...palette option 244 20 Utility Settings I O Interface Settings 247 Setting up the Oscilloscope s LAN Connection 248 To establish a LAN connection 249 Stand alone Point to Point Connection to a PC 250...

Страница 18: ...k Action Key 262 Adding an Annotation 263 21 Web Interface Accessing the Web Interface 268 Browser Web Control 269 Real Scope Remote Front Panel 270 Simple Remote Front Panel 271 Browser Based Remote...

Страница 19: ...ation 286 Licensed Options Available 286 Other Options Available 287 Upgrading to an MSO 287 Software and Firmware Updates 288 Binary Data bin Format 288 Binary Data in MATLAB 289 Binary Header Format...

Страница 20: ...Data in the Lister 316 24 I2C SPI Triggering and Serial Decode Setup for I2C Signals 317 I2C Triggering 318 I2C Serial Decode 322 Interpreting I2C Decode 323 Interpreting I2C Lister Data 324 Searchin...

Страница 21: ...nfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Interpreting UART RS232 Decode 340 UART RS232 Totalizer 341 Interpreting UART RS232 Lister Data 342 Searching for UART RS232 Data in the Lister...

Страница 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...

Страница 23: ...the Rear Panel Connectors 43 Learn the Oscilloscope Display 45 Access the Built In Quick Help 47 This chapter describes the steps you take when using the oscilloscope for the first time Inspect the P...

Страница 24: ...illoscopes User s Guide 1 Getting Started InfiniiVision 2000 X Series oscilloscope Power cord country of origin determines specific type Oscilloscope probes Two probes for 2 channel models Four probes...

Страница 25: ...Series Oscilloscopes User s Guide 25 See Also Probes and Accessories on page 285 InfiniiVision 2000 X Series oscilloscope Power cord Based on country of origin Documentation CD N2841A probes Qty 2 or...

Страница 26: ...gently remove the module from the slot 2 To install a module slide the module into the slot on the back until it is fully seated The module s spring tabs will latch into the slot keeping the module in...

Страница 27: ...be free from obstructions Unrestricted air flow is required for proper cooling Always ensure that the air intake and exhaust areas are free from obstructions The fan draws air in from the left side a...

Страница 28: ...a standby switch and is not a LINE switch Alternatively an externally installed switch or circuit breaker which is readily identifiable and is easily reached by the operator may be used as a disconnec...

Страница 29: ...ettings This places the oscilloscope in a known operating condition The major default settings are CAUTION When measuring voltages over 30 V use a 10 1 probe CAUTION Do not float the oscilloscope chas...

Страница 30: ...uld see a waveform on the oscilloscope s display similar to this Table 2 Default Configuration Settings Horizontal Normal mode 100 s div scale 0 s delay center time reference Vertical Analog Channel 1...

Страница 31: ...t the square wave is not shaped correctly as shown above perform the procedure Compensate Passive Probes on page 32 If you do not see the waveform make sure the probe is connected securely to the fron...

Страница 32: ...oscilloscope channel to which it is connected A poorly compensated probe can introduce significant measurement errors 1 Input the Probe Comp signal see Input a Waveform on page 29 2 Press Default Setu...

Страница 33: ...earn the Front Panel Controls and Connectors On the front panel key refers to any key button you can press Softkey specifically refers to the six keys that are directly below the display The legend fo...

Страница 34: ...on press again to switch power off See Power On the Oscilloscope on page 27 2 Softkeys The functions of these keys change based upon the menus shown on the display directly above the keys The Back Up...

Страница 35: ...to access the Utility Menu which lets you configure the oscilloscope s I O settings use the file explorer set preferences access the service menu and choose other options See Chapter 20 Utility Settin...

Страница 36: ...ne adjustment and select the trigger time reference point Zoom key Press the zoom key to split the oscilloscope display into Normal and Zoom sections without opening the Horizontal Menu Search key Let...

Страница 37: ...cilloscopes User s Guide 37 10 Auto Scale key When you press the AutoScale key the oscilloscope will quickly determine which channels have activity and it will turn these channels on and scale them to...

Страница 38: ...one digit channel number from 0 to 7 to D The indicates that two channels are overlaid You can rotate the upper knob to select an overlaid channel then rotate the lower knob to position it just as yo...

Страница 39: ...or High Resolution acquisition modes see Selecting the Acquisition Mode on page 161 and use segmented memory see Acquiring to Segmented Memory on page 167 The Display key lets you access the menu wher...

Страница 40: ...nformation see Chapter 3 Vertical Controls starting on page 63 17 Analog channel inputs Attach oscilloscope probes or BNC cables to these BNC connectors In the InfiniiVision 2000 X Series oscilloscope...

Страница 41: ...For more information about printing see Chapter 19 Print Screens starting on page 241 You can also use the USB port to update the oscilloscope s system software when updates are available You do not n...

Страница 42: ...ght InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started 3 Reinstall the front panel knobs Front panel overlays may be ordered from www parts keysight com using the following part...

Страница 43: ...lay 4 Channel Overlay French 75019 94324 75019 94316 German 75019 94326 75019 94318 Italian 75019 94323 75019 94331 Japanese 75019 94311 75019 94312 Korean 75019 94329 75019 94321 Polish 75019 94335 7...

Страница 44: ...to provide a larger display or to provide a display at a viewing position away from the oscilloscope The oscilloscope s built in display remains on even when an external display is connected The vide...

Страница 45: ...trigger input BNC connector See External Trigger Input on page 152 for an explanation of this feature 8 USB Device port This port is for connecting the oscilloscope to a host PC You can issue remote c...

Страница 46: ...about viewing signal detail see To adjust waveform intensity on page 113 For more information about display modes see Chapter 8 Display Settings starting on page 113 Information area The information...

Страница 47: ...p then press the Language softkey 2 Repeatedly press and release the Language softkey or rotate the Entry knob until the desired language is selected Softkey labels These labels describe softkey funct...

Страница 48: ...Vision 2000 X Series Oscilloscopes User s Guide 1 Getting Started The following languages are available English French German Italian Japanese Korean Portuguese Russian Simplified Chinese Spanish and...

Страница 49: ...adjustment setting 57 To position the time reference left center right 58 Searching for Events 58 Navigating the Time Base 59 The horizontal controls include The horizontal scale and position knobs T...

Страница 50: ...e is displayed above the Fine and Time Ref softkeys To adjust the horizontal time div scale 1 Turn the large horizontal scale sweep speed knob marked to change the horizontal time div setting Notice h...

Страница 51: ...cates how far it is from the time reference point hollow inverted triangle These reference points are indicated along the top of the display grid Figure 2 shows the trigger point with the delay time s...

Страница 52: ...ne waveform detail after the waveform has been acquired is a benefit generally associated with digital oscilloscopes Often this is simply the ability to freeze the display for the purpose of measuring...

Страница 53: ...across the screen from right to left It only operates on time base settings of 50 ms div and slower If the current time base setting is faster than the 50 ms div limit it will be set to 50 ms div when...

Страница 54: ...channel 1 and 2 position knobs Use the channel 1 and 2 volts div knobs and the channel 1 and 2 Fine softkeys to expand the signal for convenient viewing The phase difference angle can be calculated u...

Страница 55: ...cursors sin second Y first Y 1 031 1 688 37 65 degrees of phase shift NOTE Z Axis Input in XY Display Mode Blanking When you select the XY display mode the time base is turned off Channel 1 is the X a...

Страница 56: ...div window The Zoom window is a magnified portion of the normal time div window You can use Zoom to locate and horizontally expand part of the normal window for a more detailed higher resolution anal...

Страница 57: ...delay time knob is turned Negative delay values indicate you re looking at a portion of the waveform before the trigger event and positive values indicate you re looking at the waveform after the trig...

Страница 58: ...point indicator overlays the time reference indicator The time reference position sets the initial position of the trigger event within acquisition memory and on the display with delay set to 0 Turnin...

Страница 59: ...Thresholds softkey in the Search Menu to access the Measurement Threshold Menu See Measurement Thresholds on page 198 Navigating the Time Base You can use the Navigate key and controls to navigate th...

Страница 60: ...display is automatically zoomed to fit the marked row as you navigate Pressing the Scroll Lister softkey lets you use the Entry knob to scroll through data rows in the Lister display To navigate segm...

Страница 61: ...ht InfiniiVision 2000 X Series Oscilloscopes User s Guide 61 Press the navigation keys to play backward stop or play forward in time You can press the or keys multiple times to speed up the playback T...

Страница 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...

Страница 63: ...pecify bandwidth limiting 66 To change the vertical scale knob s coarse fine adjustment setting 66 To invert a waveform 67 Setting Analog Channel Probe Options 67 The vertical controls include The ver...

Страница 64: ...urn the channel on or off and to display the channel s menu When a channel is on its key is illuminated Channel Volts div Channel 1 ground level Trigger source Trigger level or threshold Channel 2 gro...

Страница 65: ...rtical expansion about the ground level of the channel however you can change this to expand about the center of the display See To choose expand about center or ground on page 253 To adjust the verti...

Страница 66: ...press the BW Limit softkey to enable or disable bandwidth limiting When bandwidth limit is on the maximum bandwidth for the channel is approximately 20 MHz For waveforms with frequencies below this t...

Страница 67: ...voltage values of the displayed waveform are inverted Invert affects how a channel is displayed However when using basic triggers the oscilloscope attempts to maintain the same trigger point by chang...

Страница 68: ...on factor must be set properly for accurate measurement results To set the probe attenuation factor 1 Press the channel key 2 Press the Probe softkey until you have selected how you want to specify th...

Страница 69: ...t Use Skew to remove cable delay errors between any two channels 1 Probe the same point with both probes 2 Press one of the probes associated channel key 3 In the Channel Menu press Probe 4 In the Cha...

Страница 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...

Страница 71: ...erformed on analog channels The resulting math waveform is displayed in light purple You can use a math function on a channel even if you choose not to display the channel on screen You can Perform an...

Страница 72: ...and re position the math waveform see To adjust the math waveform scale and offset on page 73 To perform a transform function on an arithmetic operation To perform a transform function FFT on an arith...

Страница 73: ...ath key to re size and re position the math waveform See Also Units for Math Waveforms on page 73 Units for Math Waveforms Units for each input channel can be set to Volts or Amps using the Units soft...

Страница 74: ...urce 2 values are added or subtracted point by point and the result is displayed You can use subtract to make a differential measurement or to compare two waveforms If your waveforms DC offsets are la...

Страница 75: ...tiply or divide math function the Source 1 and Source 2 values are multiplied or divided point by point and the result is displayed The divide by zero case places holes that is zero values in the outp...

Страница 76: ...T is used to compute the fast Fourier transform using analog input channels or an arithmetic operation g t FFT takes the digitized time record of the specified source and transforms it to the frequenc...

Страница 77: ...spectrum will not take up the whole screen Press the Span softkey then turn the Entry knob to set the desired frequency span of the display Center sets the FFT spectrum frequency represented at the ce...

Страница 78: ...maximum available frequency is half the FFT sample rate which is a function of the time per division setting The FFT resolution is the quotient of the sampling rate and the number of FFT points fS N T...

Страница 79: ...btained by connecting a 4 V 75 kHz square wave to channel 1 Set the horizontal scale to 50 s div vertical sensitivity to 1 V div Units div to 20 dBV Offset to 60 0 dBV Center frequency to 250 kHz freq...

Страница 80: ...e the chance that an alias will be displayed The resolution of the FFT is the effective sample rate divided by the number of points in the FFT The actual resolution of the display will not be this fin...

Страница 81: ...mpedance is set to 1 M FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50 FFT units will be displayed as dB for all other FFT sources or when a so...

Страница 82: ...e displays the spectrum The displayed waveform shows the components of the input signal above the Nyquist frequency to be mirrored aliased on the display and reflected off the right edge NOTE Nyquist...

Страница 83: ...repeats Unless there is an integral number of cycles of the sampled waveform in the record a discontinuity is created at the end of the record This is referred to as leakage In order to minimize spec...

Страница 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...

Страница 85: ...reference waveforms this happens when you press the Ref key and the LED to the left of it is lit you can use the knobs to scale and position reference waveforms There is also a skew adjustment for re...

Страница 86: ...the Clear R1 R2 softkey to clear the reference waveform location Reference waveforms are also cleared by a Factory Default or Secure Erase see Chapter 18 Save Recall Setups Screens Data starting on pa...

Страница 87: ...ion knobs to the right of the Ref key are selected for the reference waveform If the arrow to the left of the Ref key is not illuminated press the key 2 Turn the upper multiplexed knob to adjust the r...

Страница 88: ...e oscilloscope display 4 Press the Transparent softkey to enable or disable transparent information backgrounds This setting is also used for other oscilloscope information on the display like mask te...

Страница 89: ...annel 97 To change the logic threshold for digital channels 96 To display digital channels as a bus 98 Digital channel signal fidelity Probe impedance and grounding 101 This chapter describes how to u...

Страница 90: ...ble to the DIGITAL Dn D0 connector on the front panel of the mixed signal oscilloscope The digital probe cable is keyed so you can connect it only one way You do not need to power off the oscilloscope...

Страница 91: ...Series Oscilloscopes User s Guide 91 4 Connect a grabber to one of the probe leads Other probe leads are omitted from the figure for clarity 5 Connect the grabber to a node in the circuit you want to...

Страница 92: ...Digital Channels 6 For high speed signals connect a ground lead to the probe lead connect a grabber to the ground lead and attach the grabber to ground in the device under test 7 Repeat these steps u...

Страница 93: ...d displays the acquisition For digital channels each time the oscilloscope takes a sample it compares the input voltage to the logic threshold If the voltage is above the threshold the oscilloscope st...

Страница 94: ...any other key This is useful if you have unintentionally pressed the AutoScale key or do not like the settings AutoScale has selected This will return the oscilloscope to its previous settings See als...

Страница 95: ...ely toggling logic states Any channel that is turned off will be grayed out in the activity indicator To change the displayed size of the digital channels 1 Press the Digital key 2 Press the size soft...

Страница 96: ...igital channels The Digital Channel Menu is displayed above the softkeys If you want to switch the digital channels off and the Digital Channel Menu is not already displayed you must push the Digital...

Страница 97: ...There is one User softkey for each group of channels To reposition a digital channel 1 Make sure the multiplexed scale and position knobs to the right of the key are selected for digital channels If...

Страница 98: ...y to switch it on Use the Channel softkey and the Entry knob to select individual channels to be included in the bus You can rotate the Entry knob and push it or push the softkey to select channels Yo...

Страница 99: ...s To read the digital bus value at any point using the cursors 1 Turn on Cursors by pressing the Cursors key on the front panel 2 Press the cursor Mode softkey and change the mode to Hex or Binary 3 P...

Страница 100: ...anel to display the Pattern Trigger Menu and the bus values will be displayed on the right above the softkeys The dollar sign will be displayed in the bus value when the bus value cannot be displayed...

Страница 101: ...rounding Probe loading problems generally affect the device under test while probe grounding problems affect the accuracy of the data to the measurement instrument The design of the probes minimizes t...

Страница 102: ...compensating capacitor and the cable plus instrument capacitance in parallel with the stray tip capacitance to ground While this results in an input impedance specification that is an accurate model f...

Страница 103: ...to provide as much series tip resistance as possible Stray tip capacitance to ground is minimized by the proper mechanical design of the probe tip assembly This provides the maximum input impedance a...

Страница 104: ...ment will occur Sharing one probe ground with many probes forces all the current that flows into each probe to return through the same common ground inductance of the probe whose ground return is used...

Страница 105: ...digital channel group D15 D8 and D7 D0 should be attached to the ground of the device under test if any channel within the group is being used for data capture When capturing data in a noisy environme...

Страница 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...

Страница 107: ...t the oscilloscope from Auto triggering and stabilize the display You can select the trigger mode by pressing the Mode Coupling key then the Mode softkey Also the threshold voltage level must be set a...

Страница 108: ...o to www keysight com and search for the product number for example DSOX2AUTO or contact your local Keysight Technologies representative see www keysight com find contactus Lister Lister is a powerful...

Страница 109: ...ll contain all on screen serial decode information 5 Press the Scroll Lister softkey and use the Entry knob to scroll through the data Time tags in the Time column indicate the event time relative to...

Страница 110: ...try knob to scroll though data rows in the Lister display Press the Time Ref softkey and use the Entry knob to select whether the Time column in the Lister display shows times relative to the trigger...

Страница 111: ...ets you find protocol specific headers data errors etc See Searching for CAN Data in the Lister on page 309 Searching for I2C Data in the Lister on page 324 Searching for LIN Data in the Lister on pag...

Страница 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...

Страница 113: ...r various signal characteristics such as fast time div settings and low trigger rates Increasing the intensity lets you see the maximum amount of noise and infrequently occurring events Reducing the i...

Страница 114: ...114 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings Figure 13 Amplitude Modulation Shown at 100 Intensity Figure 14 Amplitude Modulation Shown at 40 Intensity...

Страница 115: ...re Waveforms softkey to perform a single shot infinite persistence A single acquisition s data is displayed with reduced intensity and it remains on the display until you clear persistence or clear th...

Страница 116: ...an also configure the Quick Action key to clear the display See Configuring the Quick Action Key on page 262 To select the grid type When the Video trigger type is selected see Video Trigger on page 1...

Страница 117: ...orresponds to the vertical sensitivity shown in the status line at the top of the display Each major horizontal division in the grid corresponds to the time div shown in the status line at the top of...

Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...

Страница 119: ...the factory default 123 You can define labels and assign them to each analog input channel or you can turn labels off to increase the waveform display area Labels can also be applied to digital channe...

Страница 120: ...ide 9 Labels 2 To turn the labels off press the Label key again To assign a predefined label to a channel 1 Press the Label key 2 Press the Channel softkey then turn the Entry knob or successively pre...

Страница 121: ...the Apply New Label softkey to assign the label to your selected channel 5 Repeat the above procedure for each predefined label you want to assign to a channel To define a new label 1 Press the Label...

Страница 122: ...a label ending in a digit such as ADDR0 or DATA0 the oscilloscope automatically increments the digit and displays the modified label in the New label field after you press the Apply New Label softkey...

Страница 123: ...duplicate labels Labels can end in any number of trailing digits As long as the base string is the same as an existing label in the library the new label will not be put in the library For example if...

Страница 124: ...es Oscilloscopes User s Guide 9 Labels NOTE Defaulting labels without erasing the default library Pressing Default Setup sets all channel labels back to the default labels but does not erase the list...

Страница 125: ...ou can also set up triggers on pulse widths patterns and video signals You can use any input channel or the External Trigger Input on page 152 BNC as the source for most trigger types Changes to the t...

Страница 126: ...ding acquisition memory into a pre trigger and post trigger buffer The position of the trigger event in acquisition memory is defined by the time reference point and the delay horizontal position sett...

Страница 127: ...ger condition is met In this mode if no triggers are occurring that is the Trig d indicator is displayed you can press Force Trigger to force a trigger and see what the input signals look like In the...

Страница 128: ...r output signal Not available when the DC or Noise waveforms are selected You can choose a channel that is turned off not displayed as the source for the edge trigger The selected trigger source is in...

Страница 129: ...ther edge mode is useful when you want to trigger on any activity of a selected source All modes operate up to the bandwidth of the oscilloscope except Either edge mode which has a limitation Either e...

Страница 130: ...analog or digital channel you want to include in the desired pattern press the Channel softkey to select the channel This is the channel source for the 0 1 X or edge condition As you press the Channe...

Страница 131: ...s in the pattern are set to don t care the oscilloscope will not trigger The rising edge or falling edge softkey sets the pattern to an edge on the selected channel Only one rising or falling edge can...

Страница 132: ...Digits softkey to set all digits to a particular value When a hex bus digit contains one or more don t care X bits and one or more bit with a value or 0 or 1 the sign will be displayed for the digit...

Страница 133: ...r analog channels turn the Trigger Level knob For digital channels press the Digital key and select Thresholds to set the threshold level The value of the trigger level or digital threshold is display...

Страница 134: ...et t 10 ns Greater than a time value For example for a positive pulse if you set t 10 ns Within a range of time values For example for a positive pulse if you set t 10 ns and t 15 ns 7 Select the qual...

Страница 135: ...d analog video signals The trigger circuitry detects the vertical and horizontal interval of the waveform and produces triggers based on the video trigger settings you have selected The oscilloscope s...

Страница 136: ...gger level because the trigger level is automatically set to the sync pulse Trigger coupling is automatically set to TV in the Trigger Mode and Coupling Menu 4 Press the sync polarity softkey to set t...

Страница 137: ...ideo Video trigger mode is set to All Lines Display Grid type is set to IRE when Standard is NTSC or mV see To select the grid type on page 116 Horizontal time division is set to 10 s div for NTSC PAL...

Страница 138: ...nt numbers per field for each video standard Video Triggering Examples The following are exercises to familiarize you with video triggering These exercises use the NTSC video standard To trigger on a...

Страница 139: ...then press the Standard softkey to select the appropriate TV standard NTSC 4 Press the Mode softkey and select the TV field of the line you want to trigger on You can choose Line Field1 Line Field2 o...

Страница 140: ...oftkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode softkey and select All Lines To trig...

Страница 141: ...ftkey and select Field1 or Field2 To trigger on all fields of the video signal To quickly and easily view transitions between fields or to find the amplitude differences between the fields use the All...

Страница 142: ...the Trigger Menu press the Trigger softkey then turn the Entry knob to select Video 3 Press the Settings softkey then press the Standard softkey to select the appropriate TV standard 4 Press the Mode...

Страница 143: ...doff softkey in the Video Trigger Menu Press the Field Holdoff softkey and use the Entry knob to adjust the holdoff in half field increments until the oscilloscope triggers on only one phase of the co...

Страница 144: ...code Options on page 107 you can enable serial trigger types To set up these triggers see CAN Triggering on page 303 I2C Triggering on page 318 LIN Triggering on page 311 Table 3 Half field holdoff ti...

Страница 145: ...Triggers 10 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 145 SPI Triggering on page 329 UART RS232 Triggering on page 337...

Страница 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...

Страница 147: ...u are probing is noisy you can set up the oscilloscope to reduce the noise in the trigger path and on the displayed waveform First stabilize the displayed waveform by removing the noise from the trigg...

Страница 148: ...Triggering and the Pre and Post Trigger Buffers After the oscilloscope starts running after pressing Run or Single or changing the trigger condition the oscilloscope first fills the pre trigger buffer...

Страница 149: ...conditions occur often enough that forced triggers are unnecessary When to Use Normal Trigger Mode The Normal trigger mode is appropriate when You only want to acquire specific events specified by th...

Страница 150: ...trigger when your waveform has a large DC offset LF low frequency Reject coupling adds a high pass filter with the 3 dB point at 50 kHz in series with the trigger waveform Low frequency reject removes...

Страница 151: ...gger the oscilloscope 1 Press the Mode Coupling key 2 In the Trigger Mode and Coupling Menu press the Noise Rej softkey to enable or disable To enable or disable trigger HF Reject HF Reject adds a 50...

Страница 152: ...g key 2 In the Trigger Mode and Coupling Menu press the Holdoff softkey then turn the Entry knob to increase or decrease the trigger holdoff time Trigger Holdoff Operating Hints The correct holdoff se...

Страница 153: ...softkey 3 In the External Trigger Menu press the Units softkey to select between Volts for a voltage probe Amps for a current probe Measurement results channel sensitivity and the trigger level will r...

Страница 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...

Страница 155: ...Stop and Single When the Run Stop key is green the oscilloscope is running that is acquiring data when trigger conditions are met To stop acquiring data press Run Stop When stopped the last acquired...

Страница 156: ...cope is stopped after running Single Single acquisitions always use the maximum memory available at least twice as much memory as acquisitions captured when running and the oscilloscope stores at leas...

Страница 157: ...order to have the signal be uniquely reconstructed without aliasing fMAX fS 2 Nyquist frequency fN folding frequency Aliasing Aliasing occurs when signals are under sampled fS 2fMAX Aliasing is the s...

Страница 158: ...n this case and it requires a system with an ideal brick wall frequency response However digital signals have frequency components above the fundamental frequency square waves are made up of sine wave...

Страница 159: ...b pdf 5989 5732EN pdf Oscilloscope Rise Time Closely related to an oscilloscope s bandwidth specification is its rise time specification Oscilloscopes with a Gaussian type frequency response have an a...

Страница 160: ...cy spectrum of fast edges where frequency components higher than fknee are insignificant in determining the shape of the signal fknee 0 5 signal rise time based on 10 90 thresholds fknee 0 4 signal ri...

Страница 161: ...ncreases and the oscilloscope s digitizer is sampling faster than is required to fill memory For example suppose an oscilloscope s digitizer has a sample period of 1 ns maximum sample rate of 1 GSa s...

Страница 162: ...In Peak Detect mode at slower time div settings when decimation would normally occur minimum and maximum value samples are kept in order to capture infrequent and narrow events at the expense of exag...

Страница 163: ...Acquisition Control 12 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 163 Figure 24 Sine With Glitch Normal Mode Figure 25 Sine With Glitch Peak Detect Mode...

Страница 164: ...ndary Press the Clear Display softkey to erase previously acquired points The display will accumulate points until Persistence is turned off 4 Characterize the glitch with Zoom mode a Press the zoom k...

Страница 165: ...hanges and how much you want to reduce the displayed noise on the signal To use the Averaging mode 1 Press the Acquire key then press the Acq Mode softkey until the Averaging mode is selected 2 Press...

Страница 166: ...olution is produced for every factor of 2 averages Random noise is reduced by for every factor of 4 averages The number of extra bits of vertical resolution is dependent on the time per division setti...

Страница 167: ...mory use the Analyze Segments feature see Infinite Persistence with Segmented Memory on page 169 to show infinite persistence across all acquired segments See also To set or clear persistence on page...

Страница 168: ...an about 1 s of inactivity consider selecting the Normal trigger mode to prevent AutoTriggering See To select the Auto or Normal trigger mode on page 148 See Also Navigating Segments on page 168 Infin...

Страница 169: ...ented memory feature is on Segmented Memory Re Arm Time After each segment fills the oscilloscope re arms and is ready to trigger in about 8 s Remember though for example if the horizontal time per di...

Страница 170: ...e sure to set the Length control to capture enough points to accurately represent the captured data When the oscilloscope is busy saving multiple segments progress is displayed in the upper right area...

Страница 171: ...rs are vertical dashed lines that adjust horizontally and can be used to measure time s frequency 1 s phase and ratio The X1 cursor is the short dashed vertical line and the X2 cursor is the long dash...

Страница 172: ...loscope and obtain a stable display 2 Press the Cursors key The Cursors box in the right side information area appears indicating that cursors are on Press the Cursors key again when you want to turn...

Страница 173: ...urned off In Hex mode the channel is interpreted as 0 if turned off 4 Press Source or X1 Source X2 Source in the Track Waveform mode then select the input source for cursor values 5 Select the cursor...

Страница 174: ...rsors softkey to set the current X1 location as 0 and the current X2 location as 100 You can press the Y Units softkey to select Base the same units used for the source waveform Ratio when selected us...

Страница 175: ...Cursors 13 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 175 Cursor Examples Figure 28 Cursors used to measure pulse widths other than middle threshold points...

Страница 176: ...t InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors Expand the display with Zoom mode then characterize the event of interest with the cursors Figure 29 Cursors measure frequency of pu...

Страница 177: ...rsors 13 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 177 Put the X1 cursor on one side of a pulse and the X2 cursor on the other side of the pulse Figure 30 Cursors track Zoom wind...

Страница 178: ...iVision 2000 X Series Oscilloscopes User s Guide 13 Cursors Press the X1 X2 linked softkey and move the cursors together to check for pulse width variations in a pulse train Figure 31 Measuring pulse...

Страница 179: ...Cursors 13 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 179 Figure 32 Moving the cursors together to check pulse width variations...

Страница 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...

Страница 181: ...reen Cursors are turned on to show the portion of the waveform being measured for the most recently selected measurement bottom most on the right side measurement are NOTE Post Acquisition Processing...

Страница 182: ...ctions or reference waveforms that are displayed are available for measurements 3 Press the Type softkey then rotate the Entry knob to select a measurement to be made NOTE If a portion of the waveform...

Страница 183: ...on some measurements 5 Press the Add Measurement softkey or push the Entry knob to display the measurement 6 To turn off measurements press the Meas key again Measurements are erased from the display...

Страница 184: ...channel waveforms as described in the following table Measurement Valid for Math FFT Valid for Digital Channels Notes Snapshot All on page 185 Amplitude on page 187 Average on page 190 Yes Full Scree...

Страница 185: ...ent type displays a popup containing a snapshot of all the single waveform measurements Rise Time on page 195 DC RMS on page 190 AC RMS on page 191 Top on page 187 Width on page 195 Yes Width on page...

Страница 186: ...using the channel Probe Units softkey See To specify the channel units on page 68 The units of math waveforms are described in Units for Math Waveforms on page 73 Peak Peak on page 186 Maximum on pag...

Страница 187: ...rsors show the values being measured Top The Top of a waveform is the mode most common value of the upper part of the waveform or if the mode is not well defined the top is the same as Maximum The Y c...

Страница 188: ...he same as Minimum The Y cursor shows the value being measured Overshoot Overshoot is distortion that follows a major edge transition expressed as a percentage of Amplitude The X cursors show which ed...

Страница 189: ...t precedes a major edge transition expressed as a percentage of Amplitude The X cursors show which edge is being measured edge closest to the trigger reference point Figure 34 Automatic Overshoot meas...

Страница 190: ...e value on all displayed data points The N Cycles measurement interval variation measures the value on an integral number of periods of the displayed signal If less than three edges are present the me...

Страница 191: ...measures the value on an integral number of periods of the displayed signal If less than three edges are present the measurement shows No edges The X cursors show the interval of the waveform being me...

Страница 192: ...as follows where x the mean N the number of measurements taken xi the ith measurement result The standard deviation is calculated as follows where the standard deviation N the number of measurements t...

Страница 193: ...94 Width on page 195 Width on page 195 Duty Cycle on page 195 Rise Time on page 195 Fall Time on page 195 Delay on page 196 Phase on page 197 Period Period is the time period of the complete waveform...

Страница 194: ...n of the waveform is being measured The Y cursor shows the middle threshold point See Also To isolate an event for frequency measurement on page 194 To isolate an event for frequency measurement The f...

Страница 195: ...hows the middle threshold point Rise Time The rise time of a signal is the time difference between the crossing of the lower threshold and the crossing of the upper threshold for a positive going edge...

Страница 196: ...nt Menu 2 Press the Source softkey then turn the Entry knob to select the first analog channel source 3 Press the Type softkey then turn the Entry knob to select Delay 4 Press the Settings softkey to...

Страница 197: ...phase shift values indicate that the rising edge of source 1 occurred after the rising edge of source 2 1 Press the Meas key to display the Measurement Menu 2 Press the Source softkey then turn the E...

Страница 198: ...ure from channel 1 to channel 2 5 Press the Back Up key to return to the Measurement Menu 6 Press the Add Measurement softkey to make the measurement The example below shows a phase measurement betwee...

Страница 199: ...units for absolute threshold for each channel is set in the channel probe menu When the Source is set to Math f t the threshold Type can only be set to Percent NOTE Changing default thresholds may ch...

Страница 200: ...per softkey then turn the Entry knob to set the upper measurement threshold value Decreasing the upper value below the set middle value will automatically decrease the middle value to be less than the...

Страница 201: ...h chosen parameters Compliance to the mask is verified point by point across the display Mask test operates on displayed analog channels it does not operate on channels that are not displayed To enabl...

Страница 202: ...alog channel is selected 7 Adjust the mask s horizontal tolerance Y and vertical tolerance X These are adjustable in graticule divisions or in absolute units volts or seconds selectable using the Unit...

Страница 203: ...is switched on when mask test is switched on then infinite persistence is switched off when mask test is switched off Troubleshooting Mask Setup If you press Create Mask and the mask appears to cover...

Страница 204: ...ayed above the softkeys Minimum Time Choose this option and then use the Test Time softkey to select how long the oscilloscope will run When the selected time has passed the oscilloscope will stop The...

Страница 205: ...mode is not available when the acquisition mode is set to Averaging Note that you can choose to Print or Save but you cannot select both at the same time All other actions may be selected at the same...

Страница 206: ...displayed Current mask name of mask Channel number date and time of Tests total number of mask tests executed Status Passing Failing or Untested Accumulated test time in hours minutes seconds and ten...

Страница 207: ...SB mass storage device and connect it to a PC 9 Open the msk file your created using a text editor such as Wordpad 10 Edit save and close the file The mask file contains the following sections Mask Fi...

Страница 208: ...s a failure Mask Violation Regions Up to 8 regions can be defined for a mask They can be numbered 1 8 They can appear in any order in the msk file The numbering of the regions must go from top to bott...

Страница 209: ...file The Y values normally voltage are normalized in the file using the following equation Ynorm Y Y1 Y where Y Y2 Y1 To convert the normalized Y values in the mask file to voltage Y Ynorm Y Y1 where...

Страница 210: ...E 06 MTES SCAL XDEL 800 000E 06 MTES SCAL Y1 359 000E 03 MTES SCAL Y2 2 35900E 00 Building a Mask File The following mask uses all eight mask regions The most difficult part of creating a mask file is...

Страница 211: ...1 500 Region Number 4 Number of vertices 5 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500 Region Number 5 Number of vertices 5 10 00 1 000 12 50 0 500 15 00 0 500 15 00 1 500 12 50 1 500...

Страница 212: ...MIN SOUR PMEM1 HARD APR AREA SCR FACT 0 FFE 0 INKS 1 PAL NONE LAY PORT SAVE FIL mask_0 SAVE IMAG AREA GRAT FACT 0 FORM NONE INKS 0 PAL COL SAVE WAV FORM NONE MTES SOUR CHAN1 ENAB 1 LOCK 1 MTES AMAS SO...

Страница 213: ...rder DSOXDVM as a stand alone item after oscilloscope purchase The DVM display is a seven segment readout like you would see on a digital voltmeter It shows the selected mode as well as the units Unit...

Страница 214: ...elect Digital Voltmeter 3 Press Features again to enable the DVM measurements 4 Press the Source softkey and turn the Entry knob to select the analog channel on which digital voltmeter DVM measurement...

Страница 215: ...ed source channel is not used in oscilloscope triggering press Auto Range to disable or enable automatic adjustment of the DVM channel s vertical scale vertical ground level position and trigger thres...

Страница 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...

Страница 217: ...u an easy way to provide input signals when testing circuitry with the oscilloscope Waveform generator settings can be saved and recalled with oscilloscope setups See Chapter 18 Save Recall Setups Scr...

Страница 218: ...ct the waveform type 3 Depending on the selected waveform type use the remaining softkeys and the Entry knob to set the waveform s characteristics CAUTION It takes the overload protection circuit abou...

Страница 219: ...Low Level softkeys to set the sine signal parameters The frequency can be adjusted from 100 mHz to 20 MHz Square Use the Frequency Frequency Fine Period Period Fine Amplitude High Level Offset Low Lev...

Страница 220: ...tor Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu press th...

Страница 221: ...u can easily set the output voltage to TTL CMOS 5 0V CMOS 3 3V CMOS 2 5V or ECL compatible Low and High levels 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope s softkey...

Страница 222: ...Rear Panel TRIG OUT Source on page 256 This is because the trigger comparator is located after the noise source To add modulation to the waveform generator output Modulation is where an original carr...

Страница 223: ...at the specified FSK rate The FSK rate specifies a digital square wave modulating signal See To set up Frequency Shift Keying Modulation FSK on page 226 To set up Amplitude Modulation AM In the Wavef...

Страница 224: ...al amplitude as the modulating signal goes from its minimum to maximum amplitude The following screen shows an AM modulation of a 100 kHz sine wave carrier signal To set up Frequency Modulation FM In...

Страница 225: ...original carrier signal frequency When the modulating signal is at its maximum amplitude the output frequency is the carrier signal frequency plus the deviation amount and when the modulating signal i...

Страница 226: ...ect Frequency Shift Keying Modulation FSK 2 Press the Hop Freq softkey and turn the Entry knob to specify the hop frequency The output frequency shifts between the original carrier frequency and this...

Страница 227: ...enerator Menu is not currently displayed on the oscilloscope s softkeys press the Wave Gen key 2 In the Waveform Generator Menu press the Settings softkey 3 In the Waveform Generator Settings Menu pre...

Страница 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...

Страница 229: ...called later You can also recall default or factory default setups Oscilloscope screen images can be saved to a USB storage device in BMP or PNG formats Acquired waveform data can be saved to a USB st...

Страница 230: ...ch displayed channel This format is also suitable for spreadsheets See To save CSV ASCII XY or BIN data files on page 232 Reference Waveform data h5 Saves waveform data in a format that can be recalle...

Страница 231: ...up files have the extension SCP These extensions appear when using the File Explorer see File Explorer on page 251 but they do not appear when using the Recall Menu To save BMP or PNG image files Imag...

Страница 232: ...o navigate to the save location See To navigate storage locations on page 236 3 Press the Settings softkey In the File Settings Menu you have these softkeys and options Setup Info when enabled setup i...

Страница 233: ...data points that will be output to the file Only displayed data points are saved The maximum number of data points depends on these things Whether acquisitions are running When stopped data comes fro...

Страница 234: ...295 Minimum and Maximum Values in CSV Files on page 296 To save Lister data files Lister data files can be saved to an external USB storage device 1 Press Save Recall Save Format then turn the Entry...

Страница 235: ...ions on page 236 6 Finally press the Press to Save softkey A message indicating whether the save was successful is displayed To save masks Mask files can be saved to one of four internal Keysight Flas...

Страница 236: ...ed location Load from when you can recall from the selected file When saving files The proposed file name is shown in the Save to file line above the softkeys To overwrite an existing file browse to t...

Страница 237: ...s on page 237 Mask msk See To recall mask files on page 238 Reference Waveform data h5 See To recall reference waveform files from a USB storage device on page 238 You can also recall setups and mask...

Страница 238: ...essful is displayed 4 If you would like to clear the display or clear the recalled mask press Clear Display or Clear Mask To recall reference waveform files from a USB storage device 1 Press the Save...

Страница 239: ...s are not changed when recalling the default setup Factory Default recalls the oscilloscope s factory default settings You must confirm the recall because there are no user settings that are left unch...

Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...

Страница 241: ...softkeys and the Press to Print softkey are ghosted not available until a printer is connected To print the oscilloscope s display 1 Connect a printer You can Connect a USB printer to either the USB p...

Страница 242: ...on page 262 See Also Adding an Annotation on page 263 To set up network printer connections When the DSOXLAN LAN VGA module is installed you can set up network printer connections A network printer i...

Страница 243: ...ter name or 192 168 1 100 650 printers printer name Path to a Windows network printer share for example server share When the Printer Address is a Windows network printer share the Modify softkey also...

Страница 244: ...ing the black background to white Invert Graticule Colors is the default mode Form Feed Select this to send a form feed command to the printer after the waveform is printed and before the setup inform...

Страница 245: ...45 Color Select this to print the screen in color The oscilloscope s print driver cannot print color images to color laser printers so the Color option is not available when connected to laser printer...

Страница 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...

Страница 247: ...Key 262 Adding an Annotation 263 This chapter explains oscilloscope utility functions I O Interface Settings The oscilloscope can be accessed and or controlled remotely via these I O interfaces USB d...

Страница 248: ...ies on page 274 Setting up the Oscilloscope s LAN Connection When the DSOXLAN LAN VGA module is installed you can place the oscilloscope on the network and set up its LAN connection Once that is done...

Страница 249: ...to the local area network LAN by inserting the LAN cable into the LAN port on the rear panel of the oscilloscope In a few moments the oscilloscope will connect to the network automatically If the osc...

Страница 250: ...escribes how to establish a point to point stand alone connection to the oscilloscope This is useful if you want to control the oscilloscope using a laptop computer or a stand alone computer 1 Press U...

Страница 251: ...SB storage device The oscilloscope s internal file system under Keysight Flash consists of 10 locations for oscilloscope setup files and four locations for mask files To use the File Explorer 1 Press...

Страница 252: ...r Delete File softkeys to take the action Pushing the Entry knob is the same as pressing the Load File softkey A file that has been deleted from a USB storage device cannot be recovered by the oscillo...

Страница 253: ...tions Preferences lets you specify oscilloscope preferences To choose expand about center or ground on page 253 To disable enable transparent backgrounds on page 254 To load the default label library...

Страница 254: ...efault label library See To reset the label library to the factory default on page 123 To set up the screen saver The oscilloscope can be configured to turn on a display screen saver when the oscillos...

Страница 255: ...disable this type of autoscale When fast debug is enabled autoscale lets you make quick visual comparisons to determine whether the signal being probed is a DC voltage ground or an active AC signal Ch...

Страница 256: ...o 1 00 PM is hour 13 The real time clock only allows selection of valid dates If a day is selected and the month or year is changed so the day is invalid the day is automatically adjusted Setting the...

Страница 257: ...See Chapter 17 Waveform Generator starting on page 217 The TRIG OUT connector also provides the User Cal signal See To perform user calibration on page 257 Performing Service Tasks The Service Menu u...

Страница 258: ...d Technology traceability is required perform the Performance Verification procedure in the Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide using traceable sources To perform use...

Страница 259: ...uide 259 For a 4 channel oscilloscope connect BNC tees to the equal length cables as shown below Then connect a BNC f to BNC f barrel connector to the tee as shown below Figure 36 User Calibration cab...

Страница 260: ...dures to verify that the oscilloscope is operating properly It is recommended you run Hardware Self Test After experiencing abnormal operation For additional information to better describe an oscillos...

Страница 261: ...tion on page 286 To display the user calibration status Pressing Utility Service User Cal Status displays the summary results of the previous User Cal and the status of probe calibrations for probes t...

Страница 262: ...nearest Keysight Technologies sales or service office for additional details Information on contacting Keysight Technologies can be found at www keysight com find contactus 1 Write the following info...

Страница 263: ...creens Data starting on page 229 Quick Recall recalls a setup mask or reference waveform Press Settings to set up the recall options See Chapter 18 Save Recall Setups Screens Data starting on page 229...

Страница 264: ...y and turn the Entry knob to select the character at the current position press this softkey to enter characters and move the cursor to the next character position press this softkey to enter characte...

Страница 265: ...hoose white red or colors that match analog channels digital channels math waveforms reference waveforms or markers 7 Press the Background softkey and turn the Entry knob to select the annotation back...

Страница 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...

Страница 267: ...er host name IP address and VISA address connect string Control the oscilloscope using the Remote Front Panel Send SCPI Standard Commands for Programmable Instrumentation remote programming commands v...

Страница 268: ...ystems Java Plug in Before you can use the web interface you must place the oscilloscope on the network and set up its LAN connection Accessing the Web Interface To access the oscilloscope s web inter...

Страница 269: ...access to The Real Scope Remote Front Panel see Real Scope Remote Front Panel on page 270 The Simple Remote Front Panel see Simple Remote Front Panel on page 271 The Browser Based Remote Front Panel...

Страница 270: ...cope using the web interface s Real Scope Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 268 2 When the oscilloscope s web interface is displayed...

Страница 271: ...ront Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 268 2 When the oscilloscope s web interface is displayed select Browser Web Control then select Simple Remo...

Страница 272: ...Panel To operate the oscilloscope using the web interface s Browser Based Remote Front Panel 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 268 2 When the oscillosco...

Страница 273: ...te programming commands to the oscilloscope via the SCPI Commands applet window 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 268 2 When the oscilloscope s web inte...

Страница 274: ...using the Keysight IO Libraries which are separate from the instrument s web interface The Keysight IO Libraries let a controller PC communicate with Keysight InfiniiVision oscilloscopes via their US...

Страница 275: ...can save setup files screen images waveform data files or mask files to your PC via the oscilloscope s web interface see Saving Files via the Web Interface on page 275 You can recall setup files refe...

Страница 276: ...lick Save The current acquisition is saved d In the File Download dialog click Save e In the Save As dialog navigate to the folder where you want to save the file then click Save Recalling Files via t...

Страница 277: ...ll Get Image To save or print the oscilloscope s display from the web interface 1 Access the oscilloscope s web interface see Accessing the Web Interface on page 268 2 When the oscilloscope s web inte...

Страница 278: ...nterface see Accessing the Web Interface on page 268 2 When the oscilloscope s web interface Welcome Page is displayed select the Identification on radio button An Identify message is displayed on the...

Страница 279: ...Web Interface 21 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 279 Install firmware upgrade files View calibration status You can select these cababilities via a drop down menu...

Страница 280: ...e via a Web browser and changing parameters Remote users can still view the Welcome screen view network status etc but they can t operate the instrument or change its setup without the password To set...

Страница 281: ...Web Interface 21 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 281 4 Enter your desired password and click Apply Changes Configure Network tab Modify Configuration...

Страница 282: ...user name is the IP address of the oscilloscope To reset the password Do one of these things to reset the password Using the keys on the front panel of the oscilloscope press Utility I O LAN Reset Us...

Страница 283: ...5 Acknowledgements 297 Product Markings and Regulatory Information 299 Specifications and Characteristics For up to date specifications and characteristics on the 2000 X Series oscilloscopes see the d...

Страница 284: ...pment Measurement category III is for measurements performed in the building installation Examples are measurements on distribution boards circuit breakers wiring including cables bus bars junction bo...

Страница 285: ...40 C Up to 45 RH up to 50 C Altitude Operating and non operating to 4 000 m 13 123 ft Overvoltage Category This product is intended to be powered by MAINS that comply to Overvoltage Category II which...

Страница 286: ...ns Available on page 287 Upgrading to an MSO on page 287 Licensed Options Available Many of the following licensed options can be easily installed without returning the oscilloscope to a Service Cente...

Страница 287: ...K Mask Limit Test Lets you create a mask and test waveforms to determine whether they comply to the mask Order DSOX2MASK mem4M Memory Upgrade It shows the total memory depth 4 Mpts interleaved Order D...

Страница 288: ...ou can use the Instrument Utilities page of the oscilloscope s web interface see Instrument Utilities on page 278 Binary Data bin Format The binary data format stores waveform data in binary format an...

Страница 289: ...of active channels multiplied by the number of segments acquired Binary Data in MATLAB Binary data from an InfiniiVision oscilloscope can be imported to The MathWorks MATLAB You can download the appr...

Страница 290: ...aveform points in the data Count A 32 bit integer that is the number of hits at each time bucket in the waveform record when the waveform was created using an acquisition mode like averaging For examp...

Страница 291: ...known 1 Volts 2 Seconds 3 Constant 4 Amps 5 dB 6 Hz Y Units A 32 bit integer that identifies the unit of measure for Y values in the acquired data The possible values are listed above under X Units Da...

Страница 292: ...e following picture shows a binary file of a single acquisition with multiple analog channels Buffer Type A 16 bit short that is the type of waveform data stored in the file 0 Unknown data 1 Normal 32...

Страница 293: ...of Waveform Buffers 1 Buffer Type 1 floating point Bytes per Point 4 Waveform Header 1 140 bytes Waveform Data Header 1 12 bytes Voltage Data 1 buffer size Waveform Header 2 140 bytes Waveform Data H...

Страница 294: ...sition on one analog channel File Header 12 bytes Waveform Header 2 140 bytes Waveform Data Header 1 12 bytes Pod 1 Timing Data buffer size Waveform Data Header 2 12 bytes Pod 2 Timing Data buffer siz...

Страница 295: ...tes Waveform Data Header 2 12 bytes Voltage Data 2 buffer size Waveform Header N 140 bytes Waveform Data Header N 12 bytes Voltage Data N buffer size Number of Waveforms N Number of Segments Number of...

Страница 296: ...ion When the oscilloscope s sample rate is 4 GSa s a sample will be taken every 250 ps If the horizontal scale is set to 10 us div there will be 100 us of data displayed because there are ten division...

Страница 297: ...Champaign Copyright 2006 2016 by The HDF Group Copyright 1998 2006 by the Board of Trustees of the University of Illinois BSD style open source jQuery Copyright 2012 jQuery Foundation and other contri...

Страница 298: ...e source distribution GNU General Public License Tabber Copyright c 2006 Patrick Fitzgerald MIT License TCLAP Copyright c 2003 Michael E Smoot MIT License time_ce Copyright C 2002 Michael Ringgaard Al...

Страница 299: ...nt are required to be separated from normal waste for disposal Reference WEEE Directive 2002 96 EC Indicates the time period during which no hazardous or toxic substance elements are expected to leak...

Страница 300: ...oise Declaration for Machines Laermangabe nach der Maschinenlaermrerordnung 3 GSGV Deutschland South Korean Certification KC mark includes the marking s identifier code which follows this format MSIP...

Страница 301: ...on AUTO or the DSOX3AUTO upgrade Setup for CAN Signals Setup consists of connecting the oscilloscope to a CAN signal using the Signals Menu to specify the signal source threshold voltage level baud ra...

Страница 302: ...baud rates from 10 kb s up to 5 Mb s or a user defined baud rate from 10 0 kb s to 4 Mb s in increments of 100 b s Fractional user defined baud rates between 4 Mb s and 5 Mb s are not allowed The defa...

Страница 303: ...s connected to an analog source channel using a differential probe Connect the probe s positive lead to the dominant low CAN signal CAN_L and connect the negative lead to the dominant high CAN signal...

Страница 304: ...t the ID and set up the number of data bytes and values Error Frame The oscilloscope will trigger on CAN active error frames All Errors The oscilloscope will trigger when any form error or active erro...

Страница 305: ...cope is stopped press the Run Stop key to acquire and decode data You can use the horizontal Zoom window for easier navigation of the decoded data See Also Interpreting CAN Decode on page 306 CAN Tota...

Страница 306: ...pears in hex digits in blue when valid or in red to indicate that the oscilloscope s hardware decode calculated a CRC that is different from the incoming CRC data stream Angled waveforms show an activ...

Страница 307: ...an overflow condition occurs the counter displays OVERFLOW The counters can be reset to zero by pressing the Reset CAN Counters softkey Types of Frames Active error frames are CAN frames in which a C...

Страница 308: ...e and no counters are incremented Interpreting CAN Lister Data In addition to the standard Time column the CAN Lister contains these columns ID frame ID Type frame type RMT remote frame or Data DLC da...

Страница 309: ...s the Bits softkey to enter the ID Remote or Data Frame ID Finds remote or data frames matching the specified ID Press the Bits softkey to select the ID Data Frame ID and Data Finds data frames matchi...

Страница 310: ...serial decode slot 7 Press the Baud Rate softkey to open the LIN Baud Rate Menu 8 Press the Baud softkey then turn the Entry knob to select the baud rate that matches match your LIN bus signal The def...

Страница 311: ...Setup for LIN Signals on page 309 LIN triggering can trigger on the rising edge at the Sync Break exit of the LIN single wire bus signal that marks the beginning of the message frame the Frame ID or t...

Страница 312: ...ame ID and data To select the frame ID value press the Frame ID softkey and use the Entry knob Note that you can enter a don t care value for the frame ID and trigger on data values only To set up the...

Страница 313: ...s selected 3 If the decode line does not appear on the display press the Serial key to turn it on 4 If the oscilloscope is stopped press the Run Stop key to acquire and decode data You can use the hor...

Страница 314: ...appears in blue if correct or red if incorrect The checksum always appears in white for LIN 2 0 Decoded text is truncated at the end of the associated frame when there is insufficient space within fr...

Страница 315: ...blue If the wakeup signal is not followed by a valid wakeup delimiter a wakeup error is detected and displayed as WUP in red Interpreting LIN Lister Data In addition to the standard Time column the L...

Страница 316: ...press the Search softkey then turn the Entry knob to select the serial slot Serial 1 or Serial 2 on which the LIN signal is being decoded 3 Press Search then select from these options ID Finds frames...

Страница 317: ...e requires Option EMBD or the DSOX3EMBD upgrade Setup for I2C Signals I2 C Inter IC bus signals setup consists of connecting the oscilloscope to the serial data SDA line and the serial clock SCL line...

Страница 318: ...level when the trigger type is set to the selected serial decode slot Data must be stable during the whole high clock cycle or it will be interpreted as a start or stop condition data transitioning w...

Страница 319: ...y Ack SCL clock bit Address with no Ack The oscilloscope triggers when the acknowledge of the selected address field is false The R W bit is ignored Restart The oscilloscope triggers when another star...

Страница 320: ...6th clock edge if all bits in the pattern match For triggering purposes a restart is treated as a start condition 10 bit Write The oscilloscope triggers on a 10 bit write frame on the 26th clock edge...

Страница 321: ...or 10 bit device address You can select from an address range of 0x00 to 0x7F 7 bit or 0x3FF 10 bit hexadecimal When triggering on a read write frame the oscilloscope will trigger after the start addr...

Страница 322: ...e 3 If the decode line does not appear on the display press the Serial key to turn it on 4 If the oscilloscope is stopped press the Run Stop key to acquire and decode data You can use the horizontal Z...

Страница 323: ...s appear in yellow along with the R character Restart addresses appear in green along with the S character Data values appear in white A indicates Ack low A indicates No Ack high Decoded text is trunc...

Страница 324: ...ress colored blue for writes yellow for reads Data data bytes Missing Ack indicated by an X highlighted in red if an error Aliased data is highlighted in pink When this happens decrease the horizontal...

Страница 325: ...if all bits in the pattern match Frame Start Address7 Write Ack Data finds a write frame on the 17th clock edge if all bits in the pattern match Frame Start Address7 Read Ack Data Ack Data2 finds a re...

Страница 326: ...lly set b Press the Threshold softkey then turn the Entry knob to select the clock signal threshold voltage level The threshold voltage level is used in decoding and it will become the trigger level w...

Страница 327: ...me the trigger level when the trigger type is set to the selected serial decode slot 7 Optional Press the MISO softkey to open the SPI Master In Slave Out Menu In the SPI Master In Slave Out Menu a Pr...

Страница 328: ...s set to CS or CS the first clock edge as defined rising or falling seen after the CS or CS signal transitions from low to high or high to low is the first clock in the serial stream Chip Select Press...

Страница 329: ...er level when the trigger type is set to the selected serial decode slot When Display Info is enabled information about the selected signal sources and their threshold voltage levels as well as a wave...

Страница 330: ...e Entry knob to set the number of bits Bits in the serial data string The number of bits in the string can be set anywhere from 4 bits to 64 bits The data values for the serial string are displayed in...

Страница 331: ...then turn the Entry knob to select the bit order most significant bit first MSB or least significant bit first LSB used when displaying data in the serial decode waveform and in the Lister 4 If the de...

Страница 332: ...frame Mid level blue lines show an idle bus The number of clocks in a frame appears in light blue above the frame to the right Decoded hexadecimal data values appear in white Decoded text is truncated...

Страница 333: ...rizontal time div setting and run again Searching for SPI Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of SPI data in the Lister You can use the N...

Страница 334: ...oftkey to open the SPI Bits Search Menu 5 In the SPI Bits Search Menu use the Words softkey to specify the number of words in the data value then use the remaining softkeys to enter the hex digit valu...

Страница 335: ...ART RS232 Serial Decode 339 UART RS232 triggering and serial decode requires Option COMP or the DSOX3COMP upgrade Setup for UART RS232 Signals To set up the oscilloscope to capture UART RS232 signals...

Страница 336: ...lowing parameters a Bits Set the number of bits in the UART RS232 words to match your device under test selectable from 5 9 bits b Parity Choose odd even or none based on your device under test c Baud...

Страница 337: ...ss the Trigger Setup softkey to open the UART RS232 Trigger Setup Menu 4 Press the Base softkey to select Hex or ASCII as the base displayed on the Data softkey in the UART RS232 Trigger Setup Menu No...

Страница 338: ...ers on a parity error based on the parity you have set in the Bus Configuration Menu 6 If you choose a trigger condition that includes Data in its description for example Rx Data then press the Data i...

Страница 339: ...d Valid ASCII characters are between 0x00 and 0x7F To display in ASCII you must select at least 7 bits in the Bus Configuration If ASCII is selected and the data exceeds 0x7F the data is displayed in...

Страница 340: ...341 Interpreting UART RS232 Lister Data on page 342 Searching for UART RS232 Data in the Lister on page 342 Interpreting UART RS232 Decode Angled waveforms show an active bus inside a packet frame Mid...

Страница 341: ...ocation of hidden data Expand the horizontal scale to allow the data to display An unknown undefined bus is shown in red A parity error will cause the associated data word to be shown in red which inc...

Страница 342: ...e the horizontal time div setting and run again Searching for UART RS232 Data in the Lister The oscilloscope s search capability lets you search for and mark certain types of UART RS232 data in the Li...

Страница 343: ...bit Finds only when the 9th alert bit is 0 The specified data byte applies to the least significant 8 bits excludes the 9th alert bit Rx X Data For use when the DUT data words are 9 bits in length inc...

Страница 344: ...344 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...

Страница 345: ...ale key 37 Auto Setup FFT 78 Auto trigger mode 148 Auto trigger indicator 149 AutoIP 249 250 automatic measurements 181 184 automatic setup 93 AutoScale preferences 255 AutoScale digital channels 93 A...

Страница 346: ...rols 38 digital channel inputs 41 digital channel menu 96 digital channels 96 digital channels AutoScale 93 digital channels enabling 287 digital channels logic threshold 96 digital channels probing 1...

Страница 347: ...47 graticule intensity 117 graticule type 116 grid intensity 117 grid type 116 ground level 64 Ground terminal 40 H Hanning FFT window 78 hardware self test 260 Help key 39 help built in 47 hex bus tr...

Страница 348: ...asurements phase 184 measurements preshoot 184 measurements time 193 measurements voltage 186 MegaZoom IV 5 mem4M 287 memory depth and sample rate 160 memory segmented 167 menu line 46 Minimum measure...

Страница 349: ...or units 174 raw acquisition record 233 Real Scope Remote Front Panel 270 rear panel connectors 43 re arm time 169 recall 263 recall files via web interface 276 recall mask files 238 recall setups 237...

Страница 350: ...ndicator 58 time re arm 169 timebase 52 times for saving data 234 Tools keys 35 Top measurement 187 totalizer CAN 307 totalizer UART rs232 341 tracking cursors 172 Traditional Chinese front panel over...

Страница 351: ...t string 267 voltage measurements 186 W warranted specifications 283 warranty 2 261 Wave Gen key 35 41 waveform generator 217 waveform generator defaults restoring 227 waveform generator expected outp...

Страница 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...

Отзывы: