06 | Keysight | Using an External Trigger to Measure Current with B2980A Series – Demo Guide
3) Rotate to select the
Measure trigger delay
and press to edit it. Then
enter
500 ms
to set the
Measure trigger delay
to
500 ms
.
4) Rotate to select the
Measure trigger period
and press to edit it. Then
enter
1 s
to set the
Measure trigger period
to
1 s
, which should be longer than the
sum of the measurement delay time and the aperture time.
5) Rotate to select the
Measure trigger source
and press to edit it.
6) Press , , to change the keys shown in Assist keys, and then
press to set the
Measure trigger source
to
Trigger In
port.
7) Press , , , and then press to open the
BNC
Configuration dialogue
.
(1) Press System Menu
(2) Press More…(1 of 2)
(3) Press I/O
(4) Press BNC…