Function Details
Trigger System
Trigger Output
B2900 can output trigger at the timing of
Trigger output
shown in
. The
trigger output terminal can be selected from the following terminals.
•
Internal bus, INT1, INT2
•
LAN port
•
Digital I/O pin, EXT1, EXT2, EXT3, EXT4, EXT5, EXT6, EXT7, EXT8,
EXT9, EXT10, EXT11, EXT12, EXT13, EXT14
The trigger output can be set by using
“Trigger Configuration dialog box” on page
. The trigger output timing and the setup parameters are shown in
.
Table 6-5
Trigger Output Timing and Setup Parameters
For using a digital I/O pin, it is necessary to set function of the pin, output trigger
polarity, type, pulse width, and the Output Trigger Timing parameter. See
Configuration dialog box” on page 4-48
“Using Digital I/O” on page 3-34
Trigger output timing
Layer
Action
Trigger
Output
Output
Trigger
Timing
Starting ARM loop
ARM
TRANS. for
transient
action or
ACQ. for
acquire
action
ON
BEFORE
Ending ARM loop
ARM
ON
AFTER
Starting TRIGGER loop
TRIGGER
ON
BEFORE
Ending TRIGGER loop
TRIGGER
ON
AFTER
Starting Transient device action
ACTION
TRANS.
ON
BEFORE
Ending Transient device action
ACTION
TRANS.
ON
AFTER
Starting Acquire device action
ACTION
ACQ.
ON
BEFORE
Ending Acquire device action
ACTION
ACQ.
ON
AFTER
Содержание B2900 Series
Страница 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Страница 3: ......
Страница 25: ...1 Getting Started...
Страница 41: ...2 Introduction...
Страница 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Страница 65: ...3 Installation...
Страница 101: ...4 Front Panel Reference...
Страница 155: ...5 Front Panel Operations...
Страница 189: ...6 Function Details...
Страница 233: ......