4-30
Keysight B2900 User’s Guide, Edition 5
Front Panel Reference
Config key group
Ranging dialog box
This dialog box provides the following parameters for setting details of the auto
ranging operation for measurements.
Ch
Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
Current Auto Ranging
Current measurement auto range operation, NORMAL,
SPEED, or RESOLUTION (RESOLN). See
.
Voltage Auto Ranging
Voltage measurement auto range operation, NORMAL,
SPEED, or RESOLUTION (RESOLN). See
.
Threshold
Sets the
rate
value of the formula shown below.
Table 4-6
Measurement Auto Range Operation
•
Basic operation
The channel automatically sets the range which provides the best resolution in
performing the measurement.
•
Upward changing operation
If measured data
value1
, the range changes up after the measurement.
value1
= measurement range
rate
/ 100
•
Downward changing operation
If measured data
value2
, the range changes down immediately.
value2
= measurement range
rate
/ 1000
Description
NORMAL
Supports basic operation and downward changing operation
described below
SPEED
Supports basic operation and upward and downward changing
operations described below
RESOLN
Supports basic operation and upward changing operation
described below
Содержание B2900 Series
Страница 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Страница 3: ......
Страница 25: ...1 Getting Started...
Страница 41: ...2 Introduction...
Страница 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Страница 65: ...3 Installation...
Страница 101: ...4 Front Panel Reference...
Страница 155: ...5 Front Panel Operations...
Страница 189: ...6 Function Details...
Страница 233: ......