Keysight B1542A User’s Guide, Edition 10
2-13
Installation
Hardware Installation
Connecting RF Probes
Only for the RF probe users. Connect the following accessories as shown in
. Use a torque wrench and an open-end wrench to fasten the SMA connectors.
Required accessories:
•
DUT cable, 1.5 m, 2 ea.
•
SMA(f)-(m)-(f) adapter, 1 ea.
•
50
Terminator, 1 ea.
•
SMA(m)-(m) adapter, 1 ea., used to connect the path to the RF probe for gate
Procedure:
1. Connect a DUT cable between a RF probe and the bias-T’s RF+DC connector or
the Drain pulse/dc switch unit’s DUT connector of the pulsed IV test system.
And set the RF probe to the appropriate place.
2. Connect the SMA adapters, terminator, and a DUT cable between a RF probe
and the divider or the Gate pulse/dc switch unit’s DUT connector of the pulsed
IV test system. And set the RF probe to the appropriate place.
Figure 2-9
RF Probe Connections
to Drain
to Gate
SMA(m)-(m)
Terminator
SMA(f)-(m)-(f)
to Bias-T (RF+DC) or
to Drain pulse/dc switch unit (DUT)
to Divider or
to Gate pulse/dc switch unit (DUT)
DUT cable, 1.5 m
DUT cable, 1.5 m
1
2
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......