Installation
Hardware Installation
Making Measurement Path for Gate
Connect the following accessories as shown in
. And fix the gate pulse/dc
switch unit to the appropriate place. Use a torque wrench and an open-end wrench to
fasten the SMA connectors.
Required accessories:
•
BNC(m)-SMA(f) adapter, 1 ea., for pulse generator
•
BNC(m)-SMA(f) adapter, precision type, 1 ea., for oscilloscope
•
BNC cable, 1 ea.
•
SMA cable, 0.3 m, 1 ea.
•
SMA cable, 1.5 m, 1 ea.
•
Triaxial cable, 1 ea.
Figure 2-7
Measurement Path for Gate
Scope
DC source monitor
Triaxial cable
Pulse generator
to SMU Force
for gate voltage
at DC test
Gate pulse/dc switch unit
SMA cable, 0.3 m
BNC cable
SMA cable, 1.5 m
to input channel
for gate (*2)
Gate Pulse Monitor
SMU
to output channel for gate pulse (*1)
(*1) Use BNC(m)-SMA(f) adapter
for pulse generator.
(*2) Use BNC(m)-SMA(f) adapter
precision type for oscilloscope.
to TRIGGER OUT
to AUX Trig In
Gate Pulse Input
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......