Configuring and Clearing Measurements
The Measure menu [Menu] > Measure allows you to view and clear the current peak hold value,
configure the measurement coupling, and specify the number of readings to average.
View and Clear IpkHold
Press [Menu] > Measure > IpkHold to display the current peak hold value. Higher peak values
overwrite previously measured lower values until power is cycled or until IpkHold is reset using Clear.
The number displayed next to ApkH is not editable; you may either press Clear or [Back] to exit. Using
[Back] does not clear the value.
SCPI:
SENSe:CURRent[:PEAK]:HOLD:CLEar
Configure Measurement Coupling
Press [Menu] > Measure > Coupling to configure measurement coupling (AC, DC, or AC+DC). The
factory default is AC+DC.
AC+DC The RMS value of the output is presented in the metering displays.
AC
The AC component of the output is presented in the metering displays.
DC
The average (DC) component of the output is presented in the metering displays.
The Meter All display is not affected by measurement coupling selection.
Specify Number of Readings to Average
Press [Menu] > Measure > Average to specify the number of readings to average for measurements.
The drop-down menu allows you to choose 1, 2, 4, 8, or 16. Press [Select] to make your selection.
SCPI:
Keysight AC6800B Series Operating and Service Guide
89
2 User Information
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...