Storing and Retrieving Instrument States
State Storage
The instrument stores the parameters shown in the table below at five-second intervals. To ensure
that settings changes are saved before power-down, wait at least five seconds after changing a
setting before turning the instrument off. Otherwise, the last settings may not be stored.
The instrument's output is always off at power up.
Stored Settings
Output mode (AC, DC, AC+DC, EXT-AC, or EXT-DC)
Output voltage range (155 V, 310 V, or AUTO)
Voltages, AC and DC
Frequency
Measured value display (RMS, PEAK, AVG, or W)
Limit values
AC
Voltage limits (lower and upper)
Current limit
Frequency limits (lower and upper)
DC
Voltage limits (lower and upper)
Current limit
Current limit operation
Save and Recall Instrument States
You can save instrument states into 10 nonvolatile state storage locations. This allows you to switch
among several configurations quickly, or to test sudden changes in output voltage or frequency.
Select [Menu] > States > SaveRecall to save or recall a state. Specify the state storage location (0
through 9) and select either Recall or Save.
If a state is recalled with the output on and includes an output mode or voltage range change, a short
alarm sounds and the state is not recalled when the output mode or voltage range switches. If this
happens, turn the output off, recall the memory, and then turn the output back on.
SCPI:
and
2 User Information
82
Keysight AC6800B Series Operating and Service Guide
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...