[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
value
>|MINimum|MAXimum[, <
low_
limit
>|MINimum|MAXimum, <
high_limit
>|MINimum|MAXimum]
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MINimum|MAXimum]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <
value
>|MINimum|MAXimum
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MINimum|MAXimum]
Sets the immediate or triggered AC voltage level, and optionally the soft limits for the immediate
voltage. The triggered level is the value that is transferred to the output when an output step is
triggered. Units are in volts (rms).
Parameter
Typical Return
AC mode 155 V range: 0.0 to 157.5 V
AC mode 310 V range: 0.0 to 315.0 V
AC+DC mode: The total voltage must be within the voltage limits, and the AC+DC peak must be between -
445 and +445 V.
+2.01
Sets the output voltage to 20 V: VOLT 20
Sets the triggered voltage to 10 V: VOLT:TRIG 10
This command takes one or three arguments. You cannot have a low or high limit without the other.
This command applies to AC and AC+DC.
Voltage units (MV, V, KV) are supported.
The
command sets this parameter to 0.
When VOLT:LIM:STAT is OFF, only individual range checking for each argument is performed. The
<
value
> must be within its own MINimum-MAXimum range considering VOLTage:RANGe and the
existing DC component setting if in ACDC coupling mode. The typical MAXimum in OUTP:COUP:AC
mode is 157.5 or 315.0 VAC, but in ACDC mode with nonzero DC components the value will be
reduced. The lower and upper soft limits must be within their MINimum-MAXimum range as defined by
the active VOLT:RANG setting. Both have the same 0 to +157.5 or +315.0 VAC limits regardless of any
operational conditions other than VOLT:RANG.
When VOLT:LIM:STAT is ON, the range checking described above is performed, and an additional
check ensures that the <
value
> does not go beyond the high and low soft limits.
The allowable range for MINimum and MAXimum varies depending on whether LIM:STAT is ON or
OFF. When LIM:STAT is OFF, the MINimum and MAXimum values are simply calculated from the
VOLT:RANG and the DC component if it is already set in ACDC mode.
3 SCPI Programming Reference
160
Keysight AC6800B Series Operating and Service Guide
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...