CURRent Subsytem
The CURRent subsystem limits the instrument's output current.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <
value
>|MINimum|MAXimum
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MINimum|MAXimum]
Sets the immediate AC current limit in amps (rms).
Parameter
Typical Return
AC6801B 0.1 to 5.2 A
AC6802B 0.2 to 10.5 A
AC6803B 0.4 to 21.0 A
AC6804B 0.8 to 42.0 A
+9.00000E-01
Sets the current limit to 900 milliamps: CURR 900 MA
Current units (UA, MA, A) are allowed.
Values exceeding the instrument's maximum value will be set to the instrument's maximum value.
The
default is the instrument's MAXimum value.
[SOURce:]CURRent:OFFSet[:IMMediate] <
value
>|MINimum|MAXimum
[SOURce:]CURRent:OFFSet[:IMMediate]? [MINimum|MAXimum]
Sets the immediate DC current limit in amps.
Parameter
Typical Return
AC6801B 0.1 to 4.2 A
AC6802B 0.2 to 8.4 A
AC6803B 0.4 to 16.8 A
AC6804B 0.8 to 33.6 A
+3.00
Set the DC current limit to 3 A: CURR:OFFS 3
Current units (UA, MA, A) are allowed.
Values exceeding the instrument's maximum value will be set to the instrument's maximum value.
The
default is the instrument's MAXimum value.
Keysight AC6800B Series Operating and Service Guide
121
3 SCPI Programming Reference
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...