Electronic Calibration Modules Reference Guide 5-15
Specifications and Characteristics
Characteristic Performance
N4431D (3.5 mm)
The characteristic performance in the following table applies to N4431D
Option 010 (3.5 mm female connectors on all ports). The data describes
performance when measuring “thru path” A-B, C-D, A–D, A-D, B–C, and B-D.
N4431D (4.3-10)
The characteristic performance in the following table applies to N4431D (if any
of the ports consist of 4.3-10 connector). The data describes performance
when measuring “thru path” A-B, C-D, A–D, A-D, B–C, and B-D.
Table 5-26 Characteristic Performance for N4431D Option 010
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz
5 GHz to 9 GHz
9 GHz to 13.5 GHz
Transmission
Tracking (±dB)
b
0.18
0.04
0.08
0.12
Load Match (dB)
35
44
41
38
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.
b. Values based on using the network analyzer N5232B 400.
Table 5-27 Characteristic Performance for N4431D (m/f 4.3–10 connectors)
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz
5 GHz to 9 GHz
9 GHz to 12 GHz
Transmission
Tracking (±dB)
b
0.18
0.05
0.09
0.12
Load Match (dB)
35
42
39
32
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.
b. Values based on using the network analyzer N5232B Option 400.
Содержание 85091D
Страница 4: ...4 ...
Страница 44: ...1 36 Electronic Calibration Modules Reference Guide General Information Safety and Regulatory Information 1 ...
Страница 66: ...2 22 Electronic Calibration Modules Reference Guide Preparing ECal for Use Operating and Safety Precautions 2 ...
Страница 82: ...3 16 Electronic Calibration Modules Reference Guide Operating ECal Module ECal Troubleshooting 3 ...
Страница 170: ...6 14 Electronic Calibration Modules Reference Guide Replaceable Parts Other ECal Accessories 6 ...
Страница 171: ......