3-2
85038A/F/M
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning
connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when
discharged. Static discharges too small to be felt can cause permanent damage. Devices such as
calibration components and devices under test (DUTs), can also carry an electrostatic charge. To prevent
damage to the test set, components, and devices:
•
always
wear a grounded wrist strap having a 1 M
resistor in series with it when handling components
and devices or when making connections to the test set.
•
always
use a grounded, conductive table mat while making connections.
•
always
wear a heel strap when working in an area with a conductive floor. If you are uncertain about
the conductivity of your floor, wear a heel strap.
•
always
ground yourself before you clean, inspect, or make a connection to a static-sensitive device or
test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
•
always
ground the center conductor of a test cable before making a connection to the analyzer test port
or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to
the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Refer to
Chapter 6 , “Replaceable Parts,”
for part numbers and instructions for ordering ESD protection
devices.
Figure 3-1
ESD Protection Setup
Содержание 85038A 7-16
Страница 1: ...Keysight Technologies 85038A F M 7 16 Calibration Kits User s and Service Guide...
Страница 2: ......
Страница 3: ...85038A F M...
Страница 6: ...85038A F M...
Страница 9: ...1 General Information...
Страница 15: ...2 Specifications...
Страница 20: ...2 6 85038A F M Specifications Electrical Specifications...
Страница 21: ...3 Use Maintenance and Care of the Devices...
Страница 29: ...4 Performance Verification...
Страница 32: ...4 4 85038A F M Performance Verification Recertification...
Страница 33: ...5 Troubleshooting...
Страница 38: ...5 6 85038A F M Troubleshooting Contacting Keysight...
Страница 39: ...6 Replaceable Parts...
Страница 41: ...85038A F M 6 3 Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85038A F M Calibration Kits...
Страница 44: ...6 6 85038A F M Replaceable Parts Introduction...
Страница 45: ...A Standard Definitions...
Страница 50: ...Index 4 85038A F M Index...