17
Keysight 16190B Performance Test Kit
Operation and Service Manual
3
Service
In this chapter, maintenance of 16190B is described.
Calibration
The performance test kit is calibrated when shipped from the factory. The
performance test kit must be calibrated at a facility that satisfies the
calibration uncertainties given in
. Keysight
Technologies’ calibration laboratories satisfy all of these calibration
uncertainties. For complete information on calibration, contact your nearest
Keysight Technologies office.
Table 3-1
Impedance Magnitude |Z| (Admittance Magnitude |Y|) Calibration
Uncertainty
1
Frequency
[MHz]
OPEN
SHORT
50
Airline with
OPEN
Airline with
SHORT
1
0.40 %
+2.5 m
0.10 %
0.16 %
0.17 %
10
0.40 %
+3.5 m
0.10 %
0.16 %
0.17 %
100
0.40 %
+14 m
0.10 %
0.16 %
0.18 %
200
0.80 %
+25 m
0.20 %
0.20 %
0.20 %
300
0.80 %
+30 m
0.20 %
0.20 %
0.20 %
500
0.80 %
+40 m
0.30 %
0.30 %
0.30 %
600
0.80 %
+40 m
0.40 %
0.50 %
0.40 %
800
0.80 %
+50 m
0.40 %
0.50 %
No Calibration
1000
0.80 %
+50 m
0.40 %
0.50 %
0.50 %
1300
0.80 %
+100 m
0.40 %
No Calibration
0.60 %