Perform calibration
In order to perform calibration more than 2 port, the M937X need to have option 551.
All the demo unit of the M937X has the option 551 installed.
In order to enable option 551 capability in your multiport configuration, you have to have at least 1 module
of the M937X which has option 551.
To check the option installed on the M937X, go to the menu bar of the M937X, then click Help > About
Network analyzer…
Click [Cal] key on the right panel
Click [Cal Wizard…] key on the menu
under “Start Cal”.
Follow instruction to perform cal.