X, P, and K 11644A
5-13
Performance Verification
System Operation Checks
9. Record the maximum peak-to-peak value in the appropriate table on
.
10. Turn off the delta marker mode.
11. Repeat this procedure from step 6, and make an S
22
measurement on the standard section. Record the
measured value in the appropriate table on
.
The S
11
and S
22
of a Thru Measurement
1. Verify that the calibration is still on.
2. Connect the standard section to ports 1 and 2 to form a thru connection.
3. To set the display:
a. Press
[MEAS]
and select
Refl: FWD S11
.
b. Press
[SCALE REF] [10] [x1]
.
c. Select
REFERENCE POSITION
and press
[10] [x1]
.
d. Select
REFERENCE VALUE
and press
[0] [x1]
.
e. Press
[MEAS]
and select
MEASURE RESTART
.
4. After one complete measurement sweep, the displayed trace should look similar to the typical
corrected trace shown in
.
5. Determine the maximum value on the trace (you can use a marker).
6. Record the maximum value in the appropriate table on
.
7. Repeat this procedure from step 3, and make an S
22
measurement. Record the measured value in the
appropriate table on
The S
21
and S
12
of a Thru Measurement
1. Verify that the calibration is still on.
2. Connect the standard section to ports 1 and 2 to form a thru connection.
3. To set the display:
a. Press
[MEAS]
and select
Trans: FWD S
21
.
b. Press
[SCALE] [.02] [x1]
.
c. Select
REFERENCE POSITION [5] [x1].
d. Select
[REFERENCE VALUE [
.2] [5] [x1]
.
e. Press
[MEAS]
and select
MEASURE RESTART
.
4. After one complete measurement sweep, the displayed trace should look similar to the typical
corrected trace shown in
.
5. If necessary, select
Press to Continue
to update the trace.
6. Determine the greatest peak-to-peak deviation (this is any positive peak to any adjacent negative peak).
You can use the markers, as described in the first test, to determine the peak-to-peak value.
7. Record the maximum peak-to-peak value in the appropriate table on
.
Содержание 11644A K
Страница 2: ......
Страница 3: ...X P and K 11644A i...
Страница 7: ...1 General Information...
Страница 12: ...1 6 X P and K 11644A General Information Preventive Maintenance...
Страница 13: ...2 Specifications...
Страница 19: ...3 User Information...
Страница 26: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition...
Страница 27: ...4 Use Maintenance and Care of the Devices...
Страница 36: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage...
Страница 37: ...5 Performance Verification...
Страница 52: ...5 16 X P and K 11644A Performance Verification System Operation Checks...
Страница 53: ...6 Troubleshooting...
Страница 57: ...7 Replaceable Parts...
Страница 58: ...7 2 X P and K 11644A Replaceable Parts...
Страница 61: ...X P and K 11644A 7 5 Replaceable Parts Introduction Figure 7 1 X Band Component Identification Sheet...
Страница 63: ...X P and K 11644A 7 7 Replaceable Parts Introduction Figure 7 2 P Band Component Identification Sheet...
Страница 65: ...X P and K 11644A 7 9 Replaceable Parts Introduction Figure 7 3 K Band Component Identification Sheet...
Страница 67: ...A Standard Definitions...