Kewtech KTD30 Скачать руководство пользователя страница 4

6

7

Continuity Test Ranges

Accuracy (Cert mode)
Ranges (Auto Range)

Tolerance (@ 20°C)

0.00 to 9.99Ω

±3% ±2 digits

10.0 to 99.9Ω

±3% ±2 digits

100 to 999Ω

±3% ±2 digits

Open Circuit Voltage

>4V, <10V

Short Circuit Current

 >200mA

Zero offset Adjust (Test Lead Null)

Typical Test Time (Cert mode) (2Ω)

<2 sec

Power

4 × AA alkaline batteries (not included)

Battery Life (BS EN 61557) > 5000 tests @ 500V test voltage

Environmental

Operating Temperature Range 

0°C to 40°C

Storage Temperature Range 

–10°C to 60°C

Size 

 

157mm × 89mm × 39mm

Weight   

400g

Insulation Test Ranges

Accuracy (Cert mode)
Test Voltage

Ranges (Auto Range)

Tolerance (@200C)

250V

0.01 to 9.99MΩ

±3% ±1 digit

10.0 to 99.9MΩ

±3% ±1 digit

100 to 199MΩ

±6% ±1 digit

500V

0.01 to 9.99MΩ

±3% ±1 digit

10.0 to 99.9MΩ

±3% ±1 digit

100 to 199MΩ

±3% ±1 digit

200 to 499MΩ

±6% ±1 digit

1000V

0.01 to 9.99MΩ

±3% ±1 digit

10.0 to 99.9MΩ

±3% ±1 digit

100 to 399MΩ

±3% ±1 digit

400 to 999MΩ

±6% ±1 digit

Output Voltage 

Voltage

Load

Output Current

Tolerance

250

250kΩ

1 mA

–0% +20%

500

500kΩ

1 mA

–0% +20%

1000

1MΩ

1 mA

–0% +20%

Short circuit current (in to 2kΩ)

<2mA

Typical Test Time (cert mode) (10MΩ)

<2 sec

Отзывы: