3. Before Making Measurements

1. Please use a AC220V power supply or Batteries (1.5 V 

Alkaline “D” size, x4) as the power supply for the PB-3103.

<When using an AC220V power supply>

 Remove the power cord from the power cord 

accessories compartment.

 Connect the plug to the socket on the main unit.

2. Insert the crushing 

handle in the correct 

position.

3.  Press the POWER Switch. The Initial screen is displayed.

4. Press the [SELECT] key to match the grain to be 

measured, and then select the measurement item.

[Note]

  When a name of application to be measured is already 

displayed, this operation is unnecessary.

ON OFF

  If characters on the display are too dark or light, 

perform contrast adjustment (see page 12).

•  Before use, leave this unit at the using site to make the 

temperature difference between the unit and enviroment within 

2°C. If the temperature between the unit and enviroment is 

not homogeneous, the temperature correction is not properly 

performed, which may lead to differences in measured values.

• How to display the unit temperature

  Turn on the POWER Switch while holding down the 

[SELECT] key in the power-off state. To hide the 

temperature, turn off the power.

<When using batteries as the power supply>

  Insert 4 size “D” batteries into the battery compartment 

on the bottom of the unit, making sure to align the positive 

 and negative

 ends correctly. In the case of using 

batteries, the power-saving mode is activated, in which 

the light amount of the LCD backlight becomes half.

  The measurement item is displayed on the screen 

in the following order every time the [SELECT] key 

is pressed: "Paddy"

"Milled Rice"

"Parboiled 

Rice"

"Parboiled Paddy"

"Tapioca Pellet"

"Tapioca 

Flour"

"TEST" 

"Paddy" and so on. When the grain 

to be measured is displayed, selection is completed.

Item changes every time the 

 key is pressed

Содержание PB-3103

Страница 1: ...Grain Tapioca Moisture TesterPB 3103 Operating Manual...

Страница 2: ......

Страница 3: ...ing Measurements 7 4 Sample Taking and Crushing 8 A Paddy Milled Rice Parboiled Rice ParboiledPaddy Tapioca Pellet 9 B Tapioca Flour 10 5 Measurements 11 6 Other functions 12 Contrast adjustment 12 Av...

Страница 4: ...cord the accessory Accessories comartment Stores the accessories to refer to the right page for the accessories Measuring section Measurement lever fixing magnet Fixes the measurement lever to prevent...

Страница 5: ...ush Batteries Power cord Operating manual 1 5 V Alkaline D size x4 Grain Tapioca Moisture Tester PB 3103 Operating Manual These accessories are stored in the accessories compartment on the top part of...

Страница 6: ...sponse Time Approximately 3 seconds Sample Temperature Automatic compensation using thermistor Compensation Format features automatic sample temperature compensation function Operating Temperature 5 4...

Страница 7: ...and enviroment within 2 C If the temperature between the unit and enviroment is not homogeneous the temperature correction is not properly performed which may lead to differences in measured values Ho...

Страница 8: ...representative of the average for the entire quantity of cereal For example samples taken from portions which are exposed to sunlight or in contact with the bottom of the container are not appropriate...

Страница 9: ...el off the spoon 3 Raise the measuring lever open the crushing compartment cover and insert the sample 4 Turn the crushing handle to crush the sample 5 Remove the sample dish and shake it lightly to f...

Страница 10: ...upplied measurement spoon with the sample and level off the spoon Remove the excess sample using a pencil or the like 2 Spoon out the flour into the sample dish and shake it lightly to flatten the flo...

Страница 11: ...ert the sample dish fully until the guide line is hid The Over Indicator oL will be displayed if the sample s moisture content exceeds this range and the Under Indicator oL will be displayed if the mo...

Страница 12: ...t will appear at the lower right of the TIMES display to indicate that the average value is being displayed AVERAGE 6 Other functions Contrast adjustment If characters on the display are too dark or l...

Страница 13: ...Raise the measurement lever and press the SELECT key again TEST disappears and the regular measurement mode resumes Note In most cases failure to display a normal value is due to the presence of dirt...

Страница 14: ...Code ASCII Printout example TEST 15 0 TEST 15 0 N 1 13 3 N 2 13 2 N 3 13 3 AV3 13 3 Printer Output Connecting the printer VZ 330 option to the connector on the rear panel of the main unit with the ca...

Страница 15: ...s of old samples adhering to the crushing rollers the new sample will be contaminated with the old sample material and accurate measurements will be impossible It is therefore necessary to clean the r...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ...s displayed on this manual may differ from the actual ones however operations and functions are not affected All efforts have been made to ensure the contents of this manual are accurate However if yo...

Отзывы: