4.4
Check the pressure using a test manometer
A minimum drive pressure of 3 bar (43 psi)
measured at the handpiece
is required to operate the
SonicFill handpiece.
3.5 bar (50 psi) is ideal
. Between 3.5 - 4.2 bar (51 - 62 psi) the drive air will
be automatically reduced within the SonicFill handpiece. The air consumption is approximately 20
- 40 NI/min. Insert the test manometer between the MULTIflex coupling and the SonicFill hand‐
piece. Adjust regulating ring on level 5.
Pressure displayed:
▪ Drive air T.R. = 3 - 4.2 bar (43 - 61 psi)
▪ Return air R.L. < 0.4 bar (6 psi)
▪ No water or spray air are needed, though.
First use
25
Содержание SonicFill
Страница 1: ...Instructions for use SonicFill Handpiece...
Страница 2: ...Manufactured for Kerr Corporation 1717 W Collins Ave Orange CA 92867 U S A...
Страница 6: ...2 Safety 2 1 1 Description of safety instructions Warning symbol Warning symbol Safety 4...
Страница 32: ...Operation 30...
Страница 57: ......
Страница 58: ...1 007 6793 kb 20120619 05 en...