WTB-BA-e-1214
10
English
Menu block
Settings
Explanation
S4
Adaptation to the environmental conditions (filter)
0
Very stable environmental conditions, weak filtering of the weighing signal
1
Normal environmental conditions, average filtering of the weighing signal
2
Unstable environmental conditions, stronger filtering of the weighing signal
3
Very unstable environmental conditions, very strong filtering of the weighing signal
4
As for setting 0, but with slower updating of the display value
5
As for setting 1, but with slower updating of the display value
6
As for setting 2, but with slower updating of the display value
7
As for setting 3, but with slower updating of the display value.
S5
Weighing unit
g
Gram
kg
Kilogram (only available if the region is set to “EU” (Europe), see menu block “CTY”)
lb
Pound (only available if the region is set to “GE” (USA), see menu block “CTY”)
oz
Ounce (only available if the region is set to “GE” (USA), see menu block “CTY”)
S6
Brightness of the display
0
Very bright
1
Bright
2
Dark
3
Very dark
DSP (S7)
Type of display
On
Weight value is only displayed when stable
Off
Weight value is displayed continuously, irrespective of whether it is stable
S8
Resolution of the scale
(only available in the user menu when “S9” = “On”)
Off
3000d
On
6000d/7500d (depending on the capacity of the scale)
S9
Access to menu item S8 from the user menu
Off
Menu item S8 is not available in the user menu (S8 is only accessible in service mode)
On
Menu item S8 is available in the user menu
S10
Extended display
Off
Normal weight display depending on the resolution of the balance (see menu item S8)
On
Only for tests: Display in d (max. scale capacity = 30,000d), keys
a
and
b
disabled
S11
Display of negative weight values
On
Display of negative weight values disabled (display shows “
oooooo
”), taring not possible
Off
Display of negative weight values enabled, taring possible (not allowed for certified balances)
S12
Speed of display when removing the load
On
Return to zero by display is fast
Off
Display returns to zero at normal speed
BAT
Type of internal power supply
CH
Rechargeable storage battery
D
Commercial battery size “D” (IEC “R20”)
GEO
Geo adjustment value
(adaptation to local gravitational conditions, see table in Chapter 5)
0 ... 31
Factory setting according to region
CAL
Scale adjustment (calibration)
No
Do not perform adjustment
Yes
Perform adjustment
(see description in Chapter 4.3)
Except where the factory settings are shown
bold
, the factory setting depends on the country in which the balance was purchased, on the model of
the balance, or on the respective weights and measures law.
Only the menu blocks shown above with a gray background are available in the user menu; the other menu blocks are only accessible in service mode.