KERN TEE 150-1 Скачать руководство пользователя страница 5

TEE-SH-e-0810 

                                   

        

5  

 

 

4.2 Function Description 

 
1.  

Load cell

 

This is the heart of the whole system. The load cell itself is arranged as a bridge.   
The resistance change of the bridge elements is proportional to the load applied on  
the load cell. Therefore, the output of the load cell is an analog signal, which is  
proportional to the load applied on the scale. 
 
 

2.  

Amplifier 

The analog signal from the load cell is very small, of the order of micro-volt.   
Hence, a linear and stable amplifier is applied to amplify the analog signal to an  
appropriate level. 
 
 

3.  

A/D  Converter 

In order for the analog signal can be input to the microprocessor, this part converts the  
analog signal to its digital equivalent. The operation of the analog to digital converter  
is using a SIGMA DELTA technique and under the control of the microprocessor. 

 
 
4.  

Microprocessor  

The microprocessor control all the functions of the scale, such as auto zero,  
A/D conversion, timings, weight calculation, display, overload indication, low battery  
indication, tare, etc…. 
 
 

5.  

Display

 

This is the part where the weight is shown out on the LCD display in digital form.   
The whole display is driven by the microprocessor. 
 
 

6.  

Power Regulator and Low Battery Detector

 

This part contains the ON/OFF power control. In order for the external power can be  
used by other parts of the scale, a regulator is used to regulate the supply.  
A low battery detector is employed to make sure that the power supply is strong  
enough for normal operation of the scale.  

 

 

7.  

Keyboard

 

The keyboard provides on user interface. [On/Zero/Off] and [Light/Cal] keys are  
employed to operate the scale. 

 
 
 
 
 
 

Содержание TEE 150-1

Страница 1: ...Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com GB Service Manual Electronic Pocket Balance Page 2 KERN TEE Version 1 0 4 2008 TEE SH e 0810 ...

Страница 2: ... Contents 1 Features 3 2 Calibration Procedure CAL 3 3 Internal Calibration Procedure 3 4 Functional Block Diagram Description 4 4 1 Functional Block Diagram 4 4 2 Function Description 5 5 Trouble Shooting 6 6 To Replace PCB 7 7 To Replace Load Cell Assembly 7 8 Schematics 8 9 Components Layout 9 ...

Страница 3: ...and repeat the procedure 3 Internal Calibration Procedure 1 Remove two screws of the rear housing and open it 2 Connect J1 on the main board by soldering the pads together 3 Place scale on a hard level surface Turn scale on Display shall show the internal counts 4 The internal counts shall fall in the range from 3 000 to 5 000 In case out of this range connect J8 left side or right side to increas...

Страница 4: ...agram Description 4 1 Functional Block Diagram 4 TEE SH e 0810 digital signal analog signal Microprocessor Power control Regulator Low battery detector Keyboard A D Converter Battery 2 x AAA Load cell Sensor Amplifier L C D ...

Страница 5: ...tion of the analog to digital converter is using a SIGMA DELTA technique and under the control of the microprocessor 4 Microprocessor The microprocessor control all the functions of the scale such as auto zero A D conversion timings weight calculation display overload indication low battery indication tare etc 5 Display This is the part where the weight is shown out on the LCD display in digital f...

Страница 6: ...der LCD Display Zero If display LO check battery 2 5V If display E check internal count Proper readout If unstable reading check load cell and environmental conditions Correct reading If not accurate perform internal calibration If cannot reach full capacity check load cell and internal zero point If always zero check internal zero point Internal calibration if necessary Normal operation ...

Страница 7: ...eck other functions such as Tare Backlight and Auto Off 7 To Replace Load Cell Assembly 1 Disassemble rear housing of scale 2 Disassemble screws fixing the load cell 3 Disconnect wires to PCB Replace the load cell with a new one Assemble screws fixing the load cell Connect wires to PCB 4 Perform internal calibration as described in section 3 5 Assemble rear housing of scale 6 Check accuracy of sca...

Страница 8: ...8 Schematics TEE 150 1 8 TEE SH e 0810 ...

Страница 9: ...9 Components Layout TEE SH e 0810 9 ...

Отзывы: