KERN TEE 150-1 Скачать руководство пользователя страница 1

KERN & Sohn GmbH 

Ziegelei 1 
D-72336 Balingen 
E-Mail: [email protected]

Tel:  +49-[0]7433-9933-0 
Fax: +49-[0]7433-9933-149 
Internet: www.kern-sohn.com 

 

 

 

 

 

 

 

GB

 

Service Manual 

Electronic Pocket Balance 

Page 2  

 

 

 

 

 
 

KERN TEE 

Version 1.0    4/2008 

 

 
 
 
 
 

 
 
 
 
 

TEE-SH-e-0810 

Содержание TEE 150-1

Страница 1: ...Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com GB Service Manual Electronic Pocket Balance Page 2 KERN TEE Version 1 0 4 2008 TEE SH e 0810 ...

Страница 2: ... Contents 1 Features 3 2 Calibration Procedure CAL 3 3 Internal Calibration Procedure 3 4 Functional Block Diagram Description 4 4 1 Functional Block Diagram 4 4 2 Function Description 5 5 Trouble Shooting 6 6 To Replace PCB 7 7 To Replace Load Cell Assembly 7 8 Schematics 8 9 Components Layout 9 ...

Страница 3: ...and repeat the procedure 3 Internal Calibration Procedure 1 Remove two screws of the rear housing and open it 2 Connect J1 on the main board by soldering the pads together 3 Place scale on a hard level surface Turn scale on Display shall show the internal counts 4 The internal counts shall fall in the range from 3 000 to 5 000 In case out of this range connect J8 left side or right side to increas...

Страница 4: ...agram Description 4 1 Functional Block Diagram 4 TEE SH e 0810 digital signal analog signal Microprocessor Power control Regulator Low battery detector Keyboard A D Converter Battery 2 x AAA Load cell Sensor Amplifier L C D ...

Страница 5: ...tion of the analog to digital converter is using a SIGMA DELTA technique and under the control of the microprocessor 4 Microprocessor The microprocessor control all the functions of the scale such as auto zero A D conversion timings weight calculation display overload indication low battery indication tare etc 5 Display This is the part where the weight is shown out on the LCD display in digital f...

Страница 6: ...der LCD Display Zero If display LO check battery 2 5V If display E check internal count Proper readout If unstable reading check load cell and environmental conditions Correct reading If not accurate perform internal calibration If cannot reach full capacity check load cell and internal zero point If always zero check internal zero point Internal calibration if necessary Normal operation ...

Страница 7: ...eck other functions such as Tare Backlight and Auto Off 7 To Replace Load Cell Assembly 1 Disassemble rear housing of scale 2 Disassemble screws fixing the load cell 3 Disconnect wires to PCB Replace the load cell with a new one Assemble screws fixing the load cell Connect wires to PCB 4 Perform internal calibration as described in section 3 5 Assemble rear housing of scale 6 Check accuracy of sca...

Страница 8: ...8 Schematics TEE 150 1 8 TEE SH e 0810 ...

Страница 9: ...9 Components Layout TEE SH e 0810 9 ...