4.2.4 F3
⇒
External calibration
Attention:
In calibrated balances the adjustment function is switch locked.
To carry out calibrations, you will have to throw the unlock switch SWA1
(See chpt. 3).
Observe stable environmental conditions. A warming up time of ca. 30 minutes is
required for stabilization. Ensure that there are no objects on the weighing plate.
•
Activate the calibrating function F3 in the menu
Weight
Basic price
Sales price
F3 WEit
CAL.
•
Press -key, the following display will appear:
Weight
Basic price
Sales price
0,000 kg
XXXXXX
(internal value)
•
Ensure that there are no objects on the weighing plate.
•
Operate menu activation key (See chpt. 4.1); stop check is carried out, then
the weight value of the calibrating weight will appear on the weight display
Weight
Basic price
Sales price
3.000
(maximum load)
Kg
(weighing unit)
XXXXXX
(internal value)
•
Carefully place adjusting weight in the centre of the weighing plate
•
Operate menu activation key (See chpt. 4.1)
Weight
Basic price
Sales price
CAL.
done
•
While the balance is counting down to zero, remove the adjusting weight.
•
The balance returns automatically into weighing mode. An error message will
appear on the display should an adjustment error occur or should the
adjustment weight be incorrect. Turn balance off, then restart it and repeat the
adjustment process.