PNJ_PNS-BA-e-1612
77
14 Error messages
Error
message
Possible cause
o-Err
Weighing range exceeded
u-Err
Insufficient preload, e. g. missing weighing pan
1-Err
Wrong adjustment weight (< 50% max)
2-Err
Divergence last external adjustment > 1%
3-Err
Weighing pan loaded during adjustment
4-Err
Divergence from last internal adjustment > 1%
A-Err
Internal adjustment automatics defective
b-Err
Check ambient conditions (static charges, vibration, etc.)
C-Err
Internal clock defective
d-Err
Damaged electronics
L-Err
Placed weight too light, e.g. for reference calculation during parts counting
or percentage determination
Err 710
Instable environmental conditions
Should other error messages occur, switch balance off and then on again. If the error