The weight value of the required adjustment weight will pop up:
Place the required calibration weight carefully in the centre of the weighing plate and
press the
key.
CAL
appears on the display, adjustment is automatically
started.
Once the adjustment has been completed,
unLoAd
will appear in the display
Take away adjustment weight
•
Return to weighing mode
Press the
button repeatedly until the "
SAVE
"? inquiry appears.
Confirm the inquiry by pressing the
button to save the changes and to return to
Weighing mode.
-or-
press the
button to return to Weighing mode without saving.