PBJ_N-SH-e-1810.docx
60
(g)Sensitivity is not
correct.
Span
calibration
has
not
been
performed correctly.
Perform span calibration.
Warm up is insufficient.
Warm up sufficiently.
Level adjustment has not been done. Adjust level correctly.
The
installation
site
has
been
changed.
Warm up sufficiently and perform span
calibration.
(UW only) Dust or dirt is stuck to
Weight(9) (built-in calibration weight).
Clean Weight(9).
(h) Large corner load error Unit assy(1) is damaged.
Perform corner load adjustment
Otherwise replace Unit assy(1).
(i) Poor measurement
repeatability / large display
value fluctuation / The
stability mark does not
readily appear.
Effects of air current / vibration /
movement of persons.
Change the balance installation site.
Remove air current / vibration /
movement of persons. Set High-stability
mode.
The sample or sample container is
charged with static electricity.
Cover the sample with metal, such as
foil.
The sample or sample container is
magnetically charged.
Demagnetize the sample and container.
Essentially large corner load error is
causing the symptoms.
Check corner load performance and
refer to (h) (Corner load error)
Essentially large drift is causing the
symptoms.
Check drift and refer to (f) (Drift)
Cable,L65(U27) is disconnected /
damaged.
Connect Cable,L65(U27) correctly /
Replace Cable,L65(U27).
Pt-Ni band(U14) is twisted/bent/badly
soldered.
Replace Pt-Ni band(U14) / Solder Pt-Ni
band(U14) correctly.
Temp sensor assy(M6) cable
is
badly soldered.
Solder Temp sensor assy(M6) cable
correctly.
The hole of Stopper plate(U11) is
contacting with the pin of Lever Assy.
Dusts exist between them.
Clean the hole of Stopper plate(U11)
and lever pin and position and lock
Stopper plate(U11) correctly.
Stopper, Lever (U12) is contacting
the
lever
end
section. Dusts
exist between them.
Clean Stopper, lever ( ) and the lever
end section and adjust the height of
Assy,Detector(U8).
Force coil assy(L1)wire is broken/
earth leak.
Replace Force coil assy(L1) .
Assy,Detector(U8) is failed.
Replace Assy,Detector(U8).
Assy,Analog(B3) is failed.
Replace Assy,Analog(B3).
Force coil assy(L1) and Magnet
assy(U6) are contacting directly or
with dusts.
Remove dusts and position Force coil
assy(L1) correctly.
Damaged EEPROM data.
Input the data with the Adjustment S/W.
Unit assy(1) is damaged.
Replace Unit assy(1).
(j) Display holds at
“CHE 0” after plugging on.
Connector(s)
is
not
properly
inserted.
Insert connectors properly.
Poor soldering of CPU (M201,
M202)
Re-solder CPU.
Poor wiring of temperature sensor
Re-solder temperature sensor cable.
Leakage of temperature sensor
Replace temperature sensor.
Содержание PBJ-N
Страница 4: ...PBJ_N SH e 1810 docx 4 ...
Страница 40: ...PBJ_N SH e 1810 docx 40 Fig 7 Fig 8 ...
Страница 47: ...47 PBJ_N SH e 1810 docx Fig 22 Fig 23 ...
Страница 55: ...55 PBJ_N SH e 1810 docx 23 Drawing Fig 31 ...
Страница 56: ...PBJ_N SH e 1810 docx 56 ...
Страница 57: ...57 PBJ_N SH e 1810 docx Fig 33 ...
Страница 58: ...PBJ_N SH e 1810 docx 58 29 Fig 34 ...