13 Functions for adaptation to ambient conditions
13.1 Stability and reaction (average value)
Exists the possibility to tune the stability of the display and the degree of reaction of
the balance to the requirements of certain applications or the installation
environment. There is a possibility of selection among five operating modes. Please
note that in general slowing down reaction times result in higher stability of the set
data handling, while speeding up reaction times have an influence on the stability
deterioration. However, the balances of PBS/PBJ type are designed to ensure both
properties i.e. fast reaction time and high stability.
13.1.1
Automatic mode
Select the menu element
22
:
The balance automatically and dynamically undertakes to optimally determine the
average value during observation of load data. If there are not any special
circumstances, this setting should always be used.
13.1.2
Filling mode
Select the menu element
23
:
This mode is suitable for weighing invariable liquid volumes. It is very susceptible to
wind and vibration.
(When the filling mode is active, the
MODE
button makes it possible to switch over
between 3 stability settings.)
13.1.3
Standard mode
Select the menu element
24
:
This mode is suitable for weighing in normal environment. The determination of the
average value is fixed and is not subject to adaptation as in the automatic mode
which is performed dynamically.
13.1.4
Antivibration mode
Select the menu element
25
:
This mode should be used if the balance is placed at a location where there are
strong vibrations, and its readings fluctuate in automatic mode.
The balance reaction is deteriorated in a result of small changes of weight.
13.1.5
Antiwind mode
Select the menu element
26
:
This mode should be used if the balance is placed at a location where it is exposed
to air currents which cause fluctuations of its readings in automatic mode.
This reaction is deteriorated even more than in the antivibration mode, but the
weighing process is more stable.