ORA_90BE_1RE-BA-e-1510
8
Important!
The ambient/room temperature and the sample temperature influence the refracto-
meter measuring result.
The scales are designed for an ambient temperature of +20 °C!
If the measurements are carried out at a temperature other than +20 °C, the results
must be correspondingly corrected.
A correction table can be found in the annex, Point 13.
7.1
Additional advice
It is important that the samples being measured are representative samples. Meas-
urements should be carried out quickly on samples that evaporate easily. The sam-
ples should be at the same temperature as the measuring instrument in order to
achieve an accurate result.