30
MBB-M-BA-e-1314
Position of the adjustment switch:
Position of the adjustment
switch
State
1. to left
Not documented
2. concentric
Adjustment position - adjustment possible
3. to right
Verification position - adjustment locked
15.3 Checking the balance verification settings
For the adjustment function, the balance must be switched over to service mode. To
achieve the effect, turn the adjustment switch to adjustment position.
In the service mode all parameters of the balance can be modified. The service
parameters may not be modified, as this could damage the balance settings.
15.3.1 Menu overview in service mode (adjustment switch in adjustment
position)
This overview is merely for checking the parameters set by the appropriate Bureau of
Standards.
Changes may only be made to the parameters for the automatic shut-off function
“
“ and the audio signal “
“.
1 2 3