1  General 

  These installation instructions cover all information required for the installation 

and start-up of the following platforms: 

KERN KFP 6V20M, KFP 6V20LM 

KERN KFP 15V20M, KFP 15V20LM 

KERN KFP 30V20SM ,KERNKFP 30V20M 

KERN KFP 60V20M, KFP 60V20LM 

KERN KFP150V20M, KFP 150V20LM 

KERN KFP300V20M 

KERN KFP600V20AM

 

2  Safety precautions 

Product safety plays an important role at KERN & Sohn. 
Non-observance of the following instructions can lead to damage to the weighing 
platform and/or injuries. 

  Before using the weighing platform read these instructions. Store these 

nstructtions for future use. 

  Take care when transporting or lifting heavy devices. 

  Only qualified personnel may install and maintain the weighing platform. 

  Disconnect the weighing terminal from the power supply before carrying out 

cleaning, installation and maintenance. 

  The weighing platform must have stabilized to room temperature before the sup-

ply voltage is switched on. 

  Do not use the weighing platform in hazardous environments. 

Содержание KFP_V20

Страница 1: ...lei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Installation instructions platform KERN KFP_V20 300 kg Version 2 2 11 2015 GB KF...

Страница 2: ...electing the site of installation 4 3 2 Package volume 4 3 3 Protect screw Exemplification 30 kg 5 3 4 Packaging return transport 6 3 5 Ambient conditions 6 3 6 Levelling 6 3 7 Connecting to the weigh...

Страница 3: ...Non observance of the following instructions can lead to damage to the weighing platform and or injuries Before using the weighing platform read these instructions Store these nstructtions for future...

Страница 4: ...he points of support At the same time it should be so stable that no vibrations arise dur ing weighing This is also to be ob served when installing the weighing platform in conveyor and similar system...

Страница 5: ...KFP_V20 300 BA e 1522 5 3 3 Protect screw Exemplification 30 kg KERN KFP KFP 6V20M KFP 6V20LM KFP 15V20M 1 2 Sealed Screws may not be loosened see figure 3 4...

Страница 6: ...ic products into liquids Observe the following ambient conditions No direct sunshine No strong draught No excessive temperature fluctuations Temperature range 10 C to 40 C 3 6 Levelling Only a weighin...

Страница 7: ...o damage The static bearing capacity i e the maximum permissible load depends on the type of load carrying position A C The maximum static bearing capacity may not be exceeded Avoid falling loads shoc...

Страница 8: ...se any hard objects to do so Do not open the weighing platform 6 Technical data 6 1 Dimensions in mm Front view Side view Model b c d f h KFP 6V20M 230 78 108 230 56 KFP 6V20LM KFP 15V20M KFP 30V20SM...

Страница 9: ...KFP_V20 300 BA e 1522 9 6 2 Technical data of the weighing cell Sensitivity 2mV V Input resistance 409 Output resistance 350 Supply voltage 10VDC OIML approval C3...

Страница 10: ...rload Protection circa kg Loadcell Capacity kg KFP 6V20 M 8 5 5 10 KFP 6V20 LM 8 5 5 10 KFP 15V20 M 23 12 30 KFP 15V20L M 23 12 30 KFP 30V20SM 46 30 50 KFP 30V20 M 46 30 50 KFP 60V20 M 85 50 100 KFP 6...

Страница 11: ...000 10 40 nLC 2 0 7 KFP 30V20SM 300x240x110 L6D D09 03 20 C3 30 0 5000 10 3000 10 40 nLC 2 0 7 KFP 30V20M 400x300x128 L6E D09 03 21 C3 50 0 6000 10 3000 10 40 nLC 2 0 7 KFP 60V20M 400x300x128 L6E D09...

Страница 12: ...176 199 49 202 172 110 1 21 42 5 80 108 1 230 x 230 cm KFP_V20 300 BA 1522...

Страница 13: ...97 1 242 257 296 6 130 236 95 42 182 26 M8 300 x 240 cm KFP_V20 300 BA 1522...

Страница 14: ...297 396 6 236 342 110 112 1 42 5 21 300 x 400 cm KFP_V20 300 BA 1522...

Страница 15: ...53 5 336 M10 30 5 457 352 109 442 400 x 500 cm KFP_V20 300 BA 1522...

Страница 16: ...424 646 6 497 587 148 11 146 25 42 5 21 500 x 650 cm KFP_V20 300 BA 1522...

Страница 17: ...516 592 782 742 187 6 51 5 30 600 x 800 cm KFP_V20 300 BA 1522...