Model FWN 300-1IP, FWN 1K0.5IP (not verifiable)
•
Press the [ON/OFF] key in order to switch off the power supply.
•
Press as well the [TARE] button and the [UNITS] button and press at the
same time the [ON/OFF] key. CAL is displayed.
Position of the UNITS key to call-up the adjustment function (not visible)
UNITS key
•
Press the [TARE] button. CAL 0 is displayed.
•
Wait for the stabilisation display and press then the [TARE] button. CAL F is
displayed.
•
For an exact calibration put the weight in the center of the weighing plate. Wait
for the stabilisation display and press then the [TARE] button.
•
The display shows END. Remove adjustment weight and the balance returns
automatically into the weighing mode.
8.8 Verification
General introduction:
According to EU directive 90/384/EEC balances must be officially verified if they are
used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes
d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard.