4 Commissioning
Attention
: Always observe chapter 2 „General Safety Instructions“!
4.1 Unpacking
SAFETY
INSTRUCTIONS
for protection against
break
Once delivered and unpacked, crane scales will not be
taken back.
The crane scales have been sealed by Messrs. KERN.
Shackles and hooks are sealed by Sella tape.
The packaging is also sealed by adhesive tape.
+
Broken seal obliges to purchase.
Thanks for your comprehension.
Your KERN Quality assurance team
Danger for the back!
The crane scales are compact and quite heavy.
Remove the scales from packaging only with the help of
a second person.
Use a lifting device such as a crane or a forklift truck.
Secure the scales that they cannot fall down when they
are lifted.
Only use original packaging for returning.
Make sure that all parts are completely present:
•
Crane scales
•
Display Unit
•
Cable screwing „Interface cable“
•
2 x mains adapter
•
Operating instructions (logbook)
4.2 Checking the original dimensions
Enter the original dimensions shown on the production data sheet in the grey
boxes of checklist chap. 10.4.
Check original dimension of crane scales; for implementation see chap. 10.3
„Regular Maintenance“
Enter all data (date, tester, results) in the first line under „Inspection before first
use“ in the checklist (see chapter 10.4)
CAUTION
If the dimensions of your first safety inspection do not match
those of KERN, the balance must not be put into operation.
In this case please contact a service partner authorised by
Messrs. KERN.