Fundamental information (general) 

2.1 

Intended use 

 
The balance you have acquired serves to determine the weighing value of the mate-
rial to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the 
material to be weighed is manually and carefully placed in the centre of the weighing 
plate. The weighing value can be read off after a stable weighing value has been 
obtained.  
 

2.2 

Inappropriate use 

 
Do not use the balance for dynamic weighing. In the event that small quantities are 
removed or added to the material to be weighed, incorrect weighing results can be 
displayed due to the “stability compensation“ in the balance. (Example: Slowly drain-
ing fluids from a container on the balance.)   
Do not leave a permanent load on the weighing plate. This can damage the measur-
ing equipment.   
Be sure to avoid impact shock and overloading the balance in excess of the pre-
scribed maximum load rating (max.), minus any possible tare weight that is already 
present.  This could cause damage to the balance.  
Never operate the balance in hazardous locations. The series design is not explo-
sion-proof.   
Structural alterations may not be made to the balance. This can lead to incorrect 
weighing results, faults concerning safety regulations as well as to destruction of the 
balance.   
The balance may only be used in compliance with the described guidelines. Varying 
areas of application/planned use must be approved by KERN in writing.   
 

2.3 

Guarantee 

 
The guarantee is not valid following 

 

- non-observation of our guidelines in the operating instructions  

- use outside the described applications  

- alteration to or opening of the device  

- mechanical damage and damage caused by media, liquids  

- natural wear and tear  

- inappropriate erection or electric installation  

- overloading of the measuring equipment  

Содержание FOB 1.5K0.5

Страница 1: ...bH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating instruction Compact balance KERN FOB Version 1 9 05 2016 GB FO...

Страница 2: ...and storage 9 4 1 Acceptance check 9 4 2 Packaging return transport 9 5 Unpacking installation and commissioning 10 5 1 Place of installation place of use 10 5 2 Installation 11 5 2 1 List of items su...

Страница 3: ...2 Weighing 18 6 3 Taring key 18 6 4 Zeroing the balance verifiable balances 18 6 5 Plus minus weighings 19 6 6 Net gross weighings 19 6 7 Auto OFF mode 19 6 8 Unit change 19 7 Maintenance upkeep disp...

Страница 4: ...2 kg 5 kg Adjustmentweight not included class 1000g M1 2000 g M3 5000 g M3 Stabilization time 2 sec Operating temperature 0 C 40 C Air humidity 25 95 non condensing Case W x D x H mm 175 x 235 x 60 W...

Страница 5: ...0 g Linearit t 4 g 10 g 20 g Adjustmentweight not included class 5 kg M3 10 kg M3 20 kg M3 Stabilization time 2 sec Operating temperature 10 C 40 C Air humidity 25 95 non condensing Case W x D x H mm...

Страница 6: ...g 1 g 2 g Linearity 1 g 2 g 4 g Adjustmentweight not included class 1000 g M2 2 000 g M3 5 000 g M2 Stabilization time 2 sec Operating temperature 0 C 40 C Air humidity 25 95 non condensing Case W x...

Страница 7: ...g 2 g Linearit t 1 5 g 3 g 6 g Adjustmentweight not included class 5 kg M1 10 kg M1 20 kg M1 Stabilization time 2 sec Operating temperature 0 C 40 C Air humidity 25 95 non condensing Case W x D x H mm...

Страница 8: ...ing equipment Be sure to avoid impact shock and overloading the balance in excess of the pre scribed maximum load rating max minus any possible tare weight that is already present This could cause dam...

Страница 9: ...ty information 3 1 Observe the information in the operating instructions Please read the operating instructions carefully before erecting and commissioning even if you already have experience with KER...

Страница 10: ...against direct draughts due to open windows and doors Avoid jarring during weighing Protect the balance against high humidity vapours and dust Do not expose the device to extreme dampness for longer p...

Страница 11: ...val by Kern 5 4 Rechargeable battery operation optional The balance may only be operated via an optional rechargeable battery If in the display lobat appears the rechargeable battery capacity is exhau...

Страница 12: ...5 7 1 Verifiable Models FOB 1K1M FOB 2K2M FOB 5K5M Switch the balance on using the key Press the adjust switch at the lower surface see chap 6 7 4 twice A numeri cal value will appear on the display...

Страница 13: ...complete adjusting 5 7 4 Check software settings for verification of a balance only models FOB 1K1M FOB 2K2M FOB 5K5M For the adjustment function the balance must be switched over to a service mode Th...

Страница 14: ...f On trAd trAtE trAn GrAv divS A_OFF CAP CAP Overload setting by hardware option Off 300 240 180 120 If the TRAD setting is OFF this must be set to ON before a verification Reset into verifiable setti...

Страница 15: ...l purposes d For the production of finished packages In case of doubt please contact your local office of weights and measures Verification information An EU qualification approval is available for th...

Страница 16: ...out the security mark Position of the security marks Balances that are subject to compulsory verification must be taken out of operation if The weighing result of the balance is outside the error limi...

Страница 17: ...17 6 Operation 6 1 Overview of display 6 1 1 Verifiable models FOB 1K1M FOB 2K2M FOB 5K5M FOB 6K2LM FOB 15K5LM FOB 30K10LM 6 1 2 Non verifiable models FOB 1 5K0 5 FOB 3K1 FOB 6K2 FOB 7 5K0 5L FOB 15K...

Страница 18: ...it for the 0 indication Place the jiffy on the weighing pan and press the key Display again shows 0 Now the weight of the jiffy is memorised internally By pressing the key after a weighing procedure 0...

Страница 19: ...is possible to control to what extent the loading area of the balance is uti lized gross i e incl weight of tare container Example Tare container 2 of max on weighing pan Tare by key the display shows...

Страница 20: ...and rub off using a clean soft cloth Loose sample residue powder can be removed carefully using a brush or hand vac uum cleaner Remove any spilt material to be weighed immediately 7 2 Maintenance upk...

Страница 21: ...ttery pack is empty The weight display changes continu ally Draught air movement Table floor vibrations The weighing plate is in contact with for eign matter Electromagnetic fields static charging cho...

Страница 22: ...ration of Conformity To view the current EC EU Declaration of Conformity go to www kern sohn com ce The scope of delivery for verified weighing balances conformity rated weighing balances includes a D...