![KERN EWJ Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/kern/ewj/ewj_operating-instructions-manual_1957232029.webp)
EWJ-BA-e-1411
29
tP
Standard printer setting
LP 50 Not documented
Acknowledge selection by
.
Eng*
Standard stetting English.
only displayed for setting „LP 50“
chi
Not documented
F4 SPD
SPd 1
Display speed
SPd 2
SPd 3
SPd 4
F5 HoUr
oFF
Automatic adjustment switched off
1 HoUr
8 HoUr
Time interval after which the automatic adjustment
starts, selectable 1, 2, 3,4, 5, 6, 7, 8 hours.
P6 bAtt
bAttoF
Mask loading status display „Recharg.battery“
bAtton
Display loading status display „Recharg.battery“
SALE m
SALE n
Sale mode no: d= 0.1g
EWJ 300-3:
d= 0.001 g
EWJ 600-2M:
d= 0.01 g
EWJ 6000-1M:
d= 0.1 g
SALE y
Sale mode yes: d= 0.01 g
EWJ 300-3:
d= 0.01 g
EWJ 600-2M:
d= 0.1 g
EWJ 6000-1M:
d= 1 g
The SALE mode is locked for balances with type approval.
To disable the access lock, destroy the seal and actuate the adjustment
switch. Position of the adjustment switch see chap. 6.11
Attention:
After destruction of the seal the balance must be re-verified by an authorised
agency and a new verification wire/seal mark fitted before it can be reused for
applications subject to verification.