32
EWJ-BA-e-1615
8.3 Overview
Factory settings are marked by *.
Menu block
Menu item
Available settings / explanation
F1 Unt
Weighing Units
Not documented
F2 bl
Display
background
illumination
EL AU*
Automatic background illumination on when
weighing pate is loaded or key pressed.
EL on
Background lighting of display is switched on
permanently
EO oFF
Display background illumination off
F3 Com
Interface
parameter
S 232
S USb
Use
to select interface: RS232 or USB
P Prt*
•
Output of a stable weighing value after pressing
of
•
Manual totalizing (menu setting „ SALE no “), see
chap. 7.9.
Press
and the weighing value will be added
to the summation memory and issued.
P Cont
Continuous data output
see chap. 9.3
P AUto
For automatic add-up see chap. 7.10.
This function is used to issue and add individual
weighing values automatically to the summation
memory on unloading of weighing scale.
(menu setting „ SALE no “)
wirel
Not documented
P ASK
Remote control instructions
Command
Function
S
Stable weighing value for the weight is sent
via the interface
W
Weighing value for the weight (stable or
unstable) is sent via the interface
T
No data are sent, the balance carries out
the tare function.
Z
No data are sent, the zero-display appears.
P
Quantity will be sent via the interface
Acknowledge selection by
.
b600
b 9600*
Baud rate
600, 1200, 2400, 4800, 9600 selectable.
Acknowledge selection by
.
Содержание EWJ 300-3