- 9 -
Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each
balance must be coordinated - in compliance with the underlying physical weighing principle -
to the existing acceleration due to gravity at its place of location (only if the balance has not
already been adjusted to the location in the factory). This adjustment process must be
carried out for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the balance periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is required for
stabilization.
Press
and
at the same time in weighing mode.
[ECF 1] is displayed.
(The adjustment process can be exited at any time with
. The balance returns then to
weighing mode.)
Confirm display of [ECF 1] with
. [CAL Z] is displayed.
Acknowledge with
. [--------] followed by [LOAd
i
] will be
displayed briefly. Then the display to input the recommended
adjustment weight is displayed
(see chap. 1 "Technical data")
The left digit flashes.
Enter the value of the adjustment weight as follows:
Switch to the next digit to the right with
.
Increase number with
.
Confirm the value entered with
.
(example)