ECB_ECE-N-BA-e-1623
13
6.4 Dosage und Zero-tracking
When the Auto-Zero-function is activated, any slight changes of the zero readout are
automatically tared.
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the “stability compensa-
tion“ in the balance. (Example: Slowly draining fluids from a container on the bal-
ance.)
Having dosages with slight changes of weight it is recommended to deactivate this
function.
With switched off
Zero-Tracking
the display however is more unstably.
Activate/deactivate Zero-Tracking
When the balance is switched off, press the
TARE
-key and keep it pressed
Do not release the
TARE
-key. Press also the
ON/OFF
-key and keep it pressed
Keep the
ON/OFF
-key pressed, however release the
TARE
-key
Press
TARE
-key anew
Keep both keys pressed until in the display “
bl
” appears
Release both keys. The balance is situated in the menu.
Press the
ON/OFF
-key until „
tr
“ appears
Confirm with the
TARE
key
Press the
ON/OFF
key to switch among the settings:
Display
Mode
„tr“
ON
Auto-Zero on
„tr“
OFF
Auto-Zero off
Confirm with the
TARE
key. Wait until the balance automatically returns to weigh-
ing mode