KERN DE120K10N Скачать руководство пользователя страница 31

DE-BA-defsi-0640

 

67

9  Data output RS 232 C 

9.1 Technical 

data 

 

ƒ

  8-bit ASCII Code 

ƒ

  1 start bit, 8 data bits, 1 stop bit, no parity bit 

ƒ

  Baud rate selectable at 1200, 2400, 4800 and 9600 Baud   

ƒ

  Miniature plug-in necessary (9 pole D-Sub) 

ƒ

  For operation with interface faultless operation is only ensured with the correct 

KERN – interface cable (max. 2m) 

9.2  Pin allocation of the balance output socket (front view) 

 

 

 

 

 

      1

 

 

                                    

9             6    

 
 
Pin 2: 

Transmit data 

Pin 3: 

Receive data 

Pin 5: 

Signal ground 

 

9.3  Explanation of the data transfer 

9.3.1 Pr 

PC 

Press the PRINT key, at stable weight the format is transferred from 

LAPR

.  

 

a.  Format for stable values for weight/quantity/percentage 
1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16 17 18
B

B B B B B B B B 0  .  0  B g  B B C

LF

 
b.  Format in case of fault  
1 2 3 4 5 6 7 8 9 10 

11 

12 

13 

14 

B B B B B B B E r  r 

 o 

 r  CR 

LF 

 

9.3.2  AU  Pr  

As soon as the weighing value is stable, the format is automatically transferred from 

LAPR

c.  Format for stable values for weight/quantity/percentage 
1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16 17 18
B

B B B B B B B B 0  .  0  B g  B B C

LF

 
d.  Format in case of fault  
1 2 3 4 5 6 7 8 9 10 

11 

12 

13 

14 

B B B B B B B E r  r  o 

 r  CR 

LF 

Содержание DE120K10N

Страница 1: ...al 48 4 2 Personnel training 48 5 Transport and storage 48 5 1 Testing upon acceptance 48 5 2 Packaging 48 6 Unpacking Setup and Commissioning 49 6 1 Installation Site Location of Use 49 6 2 Unpacking...

Страница 2: ...o tracking 62 8 4 Selection of the adjustment weight 63 8 5 Interface RS232C 64 8 5 1 Data transfer mode 64 8 5 2 Baud rate 64 8 6 Selection printed edition 65 8 7 Reset to factory setting 66 9 Data o...

Страница 3: ...arm up time 10 minutes 30 minutes 10 minutes 30 minutes Reference unit weights at piece count 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class...

Страница 4: ...minutes 30 minutes 2 hours Reference unit weights at piece count 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details for Selection of the...

Страница 5: ...minutes 2 hours 30 minutes Reference unit weights at piece count 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details for Selection of the...

Страница 6: ...inutes 10 minutes Reference unit weights at piece count 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details for Selection of the Adjustme...

Страница 7: ...30 minutes 2 hours Reference unit weights at piece count 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details for Selection of the Adjustm...

Страница 8: ...es 10 minutes 10 minutes Reference unit weights at piece count 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details for Selection of the A...

Страница 9: ...tes 10 minutes 10 minutes Reference unit weights at piece count 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details for Selection of the...

Страница 10: ...0K20NL KERN DE300K100N KERN DE300K100NL fulfill the requirements of the following CE guideline CE EMC guideline directive about electromagnetic compatibility in the version 89 336 EWG Applied harmoniz...

Страница 11: ...This may damage the measuring system Impacts and overloading exceeding the stated maximum load max of the balance minus a possibly existing tare load must be strictly avoided Balance may be damage by...

Страница 12: ...ty Precautions 4 1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning even if you are already familiar with KERN balances 4 2...

Страница 13: ...eriods of time Non permitted condensation condensation of air humidity on the appliance may occur if a cold appliance is taken to a considerably warmer environment In this case acclimatize the disconn...

Страница 14: ...Basic structure Place the balance on a horizontal and solid base refer also to 6 2 1 Installation Pull off the protection foil from the weighing plate if existing 6 2 3 1 Assembly instructions for the...

Страница 15: ...choose between the following settings 1 AF on To save the battery the balance switches off automatically 3 minutes after having finished the weighing procedure 2 AF off Switch off function deactivate...

Страница 16: ...ecommended to adjust the balance periodically in weighing operation 6 8 Adjustment The adjustment should be made with the recommended adjustment weight see chap 1 Technical data Adjustment is also pos...

Страница 17: ...tant Should the display flash or not be on 0 press the key Only now place goods onto weighing plate Take care that the weighed material does not touch the balance housing or the base mat Now the weigh...

Страница 18: ...ain Switch on balance using the key and wait for the 0 display Put tare vessel on the weighing plate and press the key 6 times until PtArE flashes on the display By actuating the key the current weigh...

Страница 19: ...ode The tare vessels can also be used for piece counting Prior to the piece count tare the tare vessel with key 7 7 Net total weighings It is useful if a mixture of several components is weighed into...

Страница 20: ...for the 0 display Press the key several times shortly The reference quantities of the counting function are passed through after that 100 is displayed Place the reference item on the weighing pan Pres...

Страница 21: ...032150747 Tael Hongkong tlh 0 02671725 Tael Taiwan tlt 0 0266666 Grain gn 15 43235835 Pennyweight dwt 0 643014931 Momme mom 0 2667 Tola tol 0 0857333381 Carat ct 5 Freely selectable factor FFA xx xx I...

Страница 22: ...X X Tael Taiwan X X X X X X X X X X X X Pennyweight X X X X X X X X X Momme X X X X X X X X X X X X Tola X X X X X X X X X X X X Freely selectable Factor X X X X X X X X X X X X Model Units DE 60K5 NL...

Страница 23: ...ay Press the key seven times in the display appears bl Confirm by pressing the key Use the key to choose between the following settings Display Adjustment Function bl on Background illumination on Con...

Страница 24: ...Function ANL Off Animal weighing function is switched off ANL 3 Weighing value calculation above 3 sec till to the value display ANL 5 Weighing value calculation above 5 sec till to the value display...

Страница 25: ...3 sec until UNIT appears By pressing the key the different menu items are called up Use the key to select a menu item Within this menu item use the key to make your choice If the key is repeatedly ac...

Страница 26: ...ically In the event that small quantities are removed or added to the material to be weighed incorrect weighing results can be displayed due to the stability compensation in the balance Example Slowly...

Страница 27: ...K1N DE15K0 5N 1000 2000 2000 4000 5000 2000 4000 4000 8000 10000 3000 6000 6000 12000 15000 DE15K5N DE24K2N DE30K1N DE36K10N DE36K10NL 5000 10000 10000 10000 10000 10000 15000 20000 20000 20000 15000...

Страница 28: ...e mode can be converted using the key Pr PC AU PC AU Pr re Cr BA Pr details chapter 9 4 AU Pr 5 The changed setting is taken over by pressing the key 6 The balance returns to weighing mode 0 0 g 8 5 2...

Страница 29: ...ion parameters using the key to scroll backward use the key After actuating the key the current state is displayed on off Change status using the or key After that the current state is saved using the...

Страница 30: ...0 display Press the key and keep it pressed until UNIT appears on the display Press the key nine times in the display appears rst Confirm selection with key The current status no is displayed Use the...

Страница 31: ...ion of the data transfer 9 3 1 Pr PC Press the PRINT key at stable weight the format is transferred from LAPR a Format for stable values for weight quantity percentage 1 2 3 4 5 6 7 8 9 10 11 12 13 14...

Страница 32: ...ASCII code After the balance having received the s w t commands it will send the following data Take into account that the following remote control commands must be sent without a subsequent CR LF s F...

Страница 33: ...ded Take into account that the output format of the balance is fixedly defined and cannot be changed The printer format is stored in the printer i e in case of a failure the printer cannot be changed...

Страница 34: ...and wipe with a dry soft cloth Loose residue sample powder can be removed carefully with a brush or manual vacuum cleaner Spilled weighing goods must be removed immediately 10 2 Service maintenance Th...

Страница 35: ...weight is permanently changing Draught air movement Table floor vibrations The weighing plate is in contact with foreign matter Electromagnetic fields static charging choose different location switch...