4.2  USING “2006” TO ENTER THE 

SERVICE

 PARAMETERS 

 

After the jumper has been placed in the position shown, apply power to the scale.   

 

The display will ask for a code number, “

Pn

” on the Weight Display.  

 

Enter the number 

2006

 when “

Pn

” is displayed and then press 

[Tare].

 

 

The displays will show the first parameter 

F1

”  “

CAL

.  To scroll through other 

parameters press the 

 key.   

 

Press the 

[Tare]

 key to enter a parameter.   

 

To exit a parameter, press the 

 key.   

 

The 

Weight

 Display window will show the parameter number and the Piece Weight 

window will show the word describing the function.   

 

When a parameter is entered by pressing the 

[Tare]

 key, the displays will guide you 

through the parameter selected and the options available. 

 

The parameters available are: 

 

F1

 

CAL

 

 

 

To enter the Calibration 

F2

 

rES

 

  Resolution 

selection, 

not valid for CPB-M units 

F3

 

CAPA

 

  Select 

Capacity 

F4

  

init           ZEro

  

Initial Zero Range 

F5

 

rE              ZEro

 

 Re-Zero 

Range 

F6

 

SCSivE   tArE

  

Successive Tare Enable 

F7

 

Ad            Count

 

Display the A/D counts 

F8

 

ZEro

 

  Set 

zero 

method. 

F9

 

Pin

 

 

 

Set new PIN number 

F10

 

LVd          ModE

 

Set low voltage detect mode 

 

 

 

 

CPB-SH-e-0721 

 

11       

 

Содержание CPB 15K0.5

Страница 1: ...2336 Balingen Frommern E Mail info kern sohn com Tel 0049 0 7433 9933 0 Fax 0049 0 7433 9933 149 Internet www kern sohn com GB Service Manual Electronic Counting Balance Page 2 KERN CPB Software rev 3...

Страница 2: ...2 USING 2006 TO ENTER THE SERVICE PARAMETERS 11 4 2 1 F1 CALIBRATION 12 4 2 1 F1 CALIBRATION 12 4 2 2 F2 RESOLUTION 12 4 2 3 F3 CAPACITY 12 4 2 4 F4 INITIAL ZERO RANGE 13 4 2 5 F5 RE ZERO RANGE 13 4...

Страница 3: ...assembly All scales available with RS 232 interface and real time clock RTC All scales have sealed keypad with colour coded membrane switches and there are three large easy to read liquid crystal typ...

Страница 4: ...ifications Interface RS 232 bi directional interface Stabilisation Time 2 Seconds typical Operating Temperature 0 C 40 C 32 F 104 F Power supply 230 VAC 50 60 Hz 120 VAC available Battery Internal rec...

Страница 5: ...able Pan rubbing against case or not installed correctly Scale not installed properly Improper connections on ADC circuit 3 Scale shows wrong weight Scale not installed correctly Check the Calibration...

Страница 6: ...h water Do not spray or immerse the scales in water Avoid air movement such as from fans or opening doors Do not place near open windows or air conditioning vents Keep the scales clean Do not stack ma...

Страница 7: ...will display the software revision number in the Weight window and count down to zero while performing a self test When ready all three displays will show zero Refer to the CPB user manual for detaile...

Страница 8: ...0000 TO ENTER THE CALIBRATION PARAMETER After the jumper has been placed in the position shown apply power to the scale The display will ask for a code number Pn on the Weight Display Enter the number...

Страница 9: ...of weight To add this value to the platform wait for stability then press the Tare key The display will show SPAn PASS if the calibration is OK Or it will show SPAn FAIL if calibration could not be co...

Страница 10: ...show dOnE if the number was accepted or FAiL if the numbers were not the same The scale returns to the parameter menu displaying F2 Pin Turn the power off remove the jumper and restart the scale The...

Страница 11: ...ht Display window will show the parameter number and the Piece Weight window will show the word describing the function When a parameter is entered by pressing the Tare key the displays will guide you...

Страница 12: ...PAn FAIL if calibration could not be completed correctly The display will show JP On Remove the calibration weight Remove the jumper and the scale will run a self test and return to normal operation 4...

Страница 13: ...parameter press the Tare key when F5 is shown The display will show the current re zero range Press the key to change the value and press Tare to accept the value NOTE Only 2 is allowed for the CPB M...

Страница 14: ...4 2 8 F8 ZERO The scale can be selected to use a normal method to determine zero Mode 1 or an alternative method Mode 2 For all CPB M scales set this option to Mode 1 To enter this parameter press th...

Страница 15: ...display will show if the LVD is on or oFF Press the key to change the value Press Tare to accept on NOTE Only LVD on is allowed for the CPB M scales Turn the power off remove the jumper and restart th...

Страница 16: ...5 0 WIRING DIAGRAM CPB SH e 0721 16...

Страница 17: ...6 0 MECHANICAL ASSEMBLY CPB SH e 0721 17...

Страница 18: ...7 0 FRONT DISPLAY SCHEMATIC CPB SH e 0721 18...

Страница 19: ...7 1 FRONT DISPLAY ASSEMBLY CPB SH e 0721 19...

Страница 20: ...8 0 MAIN PCB SCHEMATIC CPB SH e 0721 20...

Страница 21: ...8 1 MAIN PCB ASSEMBLY CPB SH e 0721 21...

Страница 22: ...ies number e g WP0702716 Label on the left side of the balance Program number e g U3 06A appears on the display when switching on the balance As precise a description of the defect as possible Weighin...

Отзывы: