14 Data output RS 232 “P4 Print“
You can print weighing data automatically via the RS 232C interface or manually
pressing the
button via the interface according to the setting in the menu.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between
the weighing system and the printer.
•
Use a suitable cable to connect the display unit to the interface of the printer.
Faultless operation requires an adequate KERN interface cable.
•
Communication parameters (baud rate, bits and parity) of display unit and
printer must match.
14.1 Technical data
Connection
9 pin d-subminiature bushing
Pin 2: Receive data
Pin 3: Transmit data
Pin 5: Signal ground
Baud rate
2400/4800/9600/19200 selectable
Parity
8 bits, no parity / 1 stop bit
Transfer
modes
•
Manually after pressing the
key
•
Automatically according to stability display
•
Continuously, according to setting
•
On request by external device (For remote control commands,
see chap. 14.6)