KERN Optics ODC-2 Скачать руководство пользователя страница 19

ODC-2-BA-e-1910 

 

  18 

  Adjust the scale on the image to match the length of the physical scale 

ruler 
You can move the scale or alter the length using the touch function. 

 

In the example shown, a scale with a fine division of 0.01 mm (10 µm) is 
used. The total length is 1000 µm 
 
 
 

  Enter  an  appropriate  name  for  the 

calibration. 
Usually  the  lens  magnification  set  for  the 
microscope is selected.  
In the example here: 4x. 

 

It  is  also  vital  to  specify  the  length  of  the 
part as established using the scale, as well 
as  the  preferred  measuring  unit  for  the 
measurement.  
In the example here: 1000 µm. 
 

 
 
 
 
 
 
 

 

Click  on  “Calculate”  in  order  to  save 
and close the calibration.  
This  procedure  must  be  carried  out 
again for other lens magnifications.  

 
 

 

Содержание ODC-2

Страница 1: ...Sohn GmbH Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User Instructions Tablet camera for microscopes KERN ODC 2 ODC 24...

Страница 2: ......

Страница 3: ...Nomenclature 5 4 Technical data 9 5 Operation 11 5 1 Unpacking and setting up 11 5 2 Start microscope software S EYE 13 5 3 User interface and functions of the S EYE microscope software 14 5 3 1 Effe...

Страница 4: ...ct sunlight for long periods of time Where possible the tablet should not be used in humid environments Using it in an environment which is always dry will automatically increase the service life 1 2...

Страница 5: ...ODC 2 BA e 1910 4 2 Scope of supply Tablet camera Power adapter User Instructions...

Страница 6: ...5 ODC 2 BA e 1910 3 Nomenclature...

Страница 7: ...ODC 2 BA e 1910 6...

Страница 8: ...7 ODC 2 BA e 1910...

Страница 9: ...mber Description 1 Tablet Touch display 7 USB port PC mouse connection 2 Camera housing 8 Connection socket for headphones 3 Camera sensor 9 Loud speaker 4 Power connector 10 Micro SD slot 5 Microphon...

Страница 10: ...ensor 1 2 5 CMOS Frames per second fps 30 fps 1024x768 Pixel size 2 2 m x 2 2 m Photo format JPEG Input voltage 12 V DC no battery operation Operating system Android 5 1 Interfaces WLAN USB 2 0 Micro...

Страница 11: ...ODC 2 BA e 1910 10...

Страница 12: ...let 3 Undo the screws on the cover on the back of the camera and remove the protective film underneath 4 Fit the tablet camera to the trinocular connection of the microscope To do this you need a C mo...

Страница 13: ...ODC 2 BA e 1910 12 Tablet camera C mount adapter Microscope...

Страница 14: ...a permanent power supply Battery operation is not possible 2 The S EYE microscope software usually starts automatically If this does not happen then this App from the desktop of the tablet can be star...

Страница 15: ...s the very best digital sample analysis As well as live image transfer the following tools are also included can be selected on the righthand edge of the scree 1 Effect Setting the camera parameters 2...

Страница 16: ...s adjusted manually 2 Auto WB white balance Auto mode Colour temperature adjusts automatically Manual mode Colour temperature is adjusted manually 3 Further image processing Brightness Contrast Satura...

Страница 17: ...recorded All images and videos are saved and can be recalled using the Playback button In order to open the required image or video you need to double click You can specify the default storage locatio...

Страница 18: ...ification settings of the microscope and the image characteristics of live transfer on the tablet camera Calibration 1 Place the object holder with integrated scale onto the stage of the microscope 2...

Страница 19: ...ngth is 1000 m Enter an appropriate name for the calibration Usually the lens magnification set for the microscope is selected In the example here 4x It is also vital to specify the length of the part...

Страница 20: ...line Adding a cross line Multiple x or y axes possible Angle Measuring an angle Double circle Measuring the distance between two circle centres Perpendicular Measuring the distance between a line and...

Страница 21: ...Measure program category It is selected when you see white background with black text 2 Click on the symbol for line measurement Then the measuring line along with the measurements will be displayed...

Страница 22: ...order to be able to use this tool it must first be activated in the Setting program category Tick the box for Enable built in image viewer As soon as an image is opened in the Playback program categor...

Страница 23: ...ODC 2 BA e 1910 22 Four different instruments for image processing or image analysis can be implement with this program 1 Greyscale GREYSCALE function 2 Contrast and brightness CONTRAST function...

Страница 24: ...23 ODC 2 BA e 1910 3 Threshold THRESHOLD function 4 Particle analysis PARTICLES function...

Страница 25: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...