3-10
MST-MHT SERIES 071320
3.9.2
+5 VOLTS REFERENCE ADJUST (R49)
This procedure establishes the +5V reference voltage used by the analog circuits.
1. Disconnect the load and turn on the UUT. Proceed to step 3 if VOLTAGE MODE indicator is
on.
2. If the CURRENT MODE indicator is on, issue a command from the host computer to set the
UUT to Voltage mode. Verify that VOLTAGE MODE indicator is on.
3. Issue a command from the host computer to reset the UUT. This causes the UUT to be pro-
grammed to zero volts.
4. Connect the voltmeter to test connector J5, pins 1 (REF) and 6 (GND) (see Figure 2-2) and
verify meter reads between 5.00000V ± 500
V volts. If meter reading is outside these limits,
adjust potentiometer R49 for reading of +5.00000V ± 50
V.
3.9.3
OUTPUT VOLTAGE (E
O
) ZERO ADJUST (R11)
This procedure adjusts the offset of the voltage comparison amplifier to be zero for a pro-
grammed output voltage of zero.
1. Configure DC OUTPUT connector as shown in Figure 3-1 and open S2.
2. With the load disconnected, issue commands from the host computer to enable the output
(output relays closed), e.g.
VOLT 0;CURR 5;:OUTP ON
. Verify OUTPUT ENABLED and
VOLTAGE MODE indicators are on.
3. Connect DVM to DC output connector (see Figure 2-3) pin 5, Sense 1 (+) and pin 2, Sense 2
(–) and verify that output is 0V ± 500
V.
4. If the measured output is outside these limits adjust potentiometer R11 for reading of
0.00000V ± 50
V.
3.9.4
CURRENT SENSE ZERO ADJUST (R50)
This procedure adjusts the current sensing amplifier to zero with no current flowing through the
current sensing resistor.
1. Issue commands (
*RST;:OUTP ON
) from the host computer to reset the UUT.
2. Verify VOLTAGE MODE indicator is on.
3. Connect voltmeter to test connector J5, pins 4 (CSNS - Current Sense) and 6 (GND) and
verify meter reads 0.00000V ± 500
V. If meter reading is outside these limits, adjust potenti-
ometer R50 for 0.00000V ± 50
V.
Содержание MST 100-2MHT
Страница 2: ......
Страница 8: ...iv MST SERIES 071320 FIGURE 1 1 MST POWER SUPPLY...
Страница 13: ...MST MHT SERIES 071320 1 5 FIGURE 1 1 MST MHT POWER SUPPLY MECHANICAL OUTLINE DRAWING...
Страница 16: ......
Страница 30: ......
Страница 39: ...MST MHT SERIES 071320 3 9 FIGURE 3 1 CALIBRATION TEST SET UP O2 S2 CS GND S1 O1...