PDS20-18/PDS36-10/PDS60-6
20
ADJUSTMENT
4. Press the MEMORY key. (The PRESET 1 to PRESET 3 LEDs goon.)
5. Press the PRESET 2key. (The PRESET 1 LED goes on.)
2-4. Current Monitor Offset
1. Press the CURRENT key. (The CURRENT LED goes on.)
2. Press the PRESET 3 key. (The PRESET 3 LED goes on.)
3. Wait until operation becomes stable. (For about ten seconds.)
4. Press the MEMORY key. (The PRESET to
1
PRESET 3 LEDs go on.)
5. Press the PRESET 3 key. (The PRESET 3 LED goes on.)
2-5. Current Monitor Gain
1. Press the CURRENT key. (The CURRENT LED goes on.)
2. Press the KEYLOCK/LOCAL key.
3. Wait until operation becomes stable. (For about ten seconds)
4. Press the MEMORY key. (The PRESET 1 to PRESET 3 LEDs go on.)
5. Press the KEYLOCK/LOCAL key.
6. Press the OUTPUT key. (The OUTPUT
LED goeson.)
3. Over
Voltage
Protect
Before
starting OVP adjustment, disconnect the shunt resister from the output terminals.
1. Rotate the OVP control fully counterclockwise to the minimum position.
2. Press the OUTPUT
key. (The OUTPUT
LED goes on.)
3. Press the OVP key.
4. The 7-segment
LED begins to count. Wait until the value stops changing.
5. OVP adjustment is completed.
PDS-20-18(K)
1,1(
98.12.3
17:07