SELECTING MEASUREMENT MODE
The KS-TB-100 has two measurement modes. These are single and continuous.
1. Press and hold the CAL key to enter setup mode.
2. Press
◄ or ► key until the display show ‘test/P-07
3.
Press ENTER key to confirm, the meter shows “1/tESt” indicating that the single measurement mode is selected.
4. Press
◄ or ► key, the meter shows “2/tESt” indicating that the continuous measurement mode is selected.
5. Select the desired measurement mode, press ENTER key to confirm. The meter returns to measurement mode.
CALIBRATION
GUIDELINES FOR MEASUREMENT AND CALIBRATION
1. Do not hold the meter in the hand during measurement and calibration.
2. Glass vials and caps must be cleaned thoroughly with deionised water after each measurement. Minor residuals can
cause errors.
3. The outside of the vial must be clean and dry, before starting the test. Wipe the vial with a lint-free cloth to remove
fingerprints or waterdrops.
4. If the vial has minor scratches or scuffs, add a few drops of silicone oil to outside of the vial. Wipe with lint-free cloth.
5. For the low turbidity samples (0 to 200 NTU), we recommend that you use the same vial to perform the
measurement or calibration.