User Manual
KS-DO-10 |Pocket Dissolved Oxygen Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
[email protected]
www.kensingtonscientific.com.au
SETUP MENU
The KS-SAL-
20 contains integrated setup menu that allows you to customize your tester’s features to fit your
measurement requirements.
Parameter
Description
Options
Description
SALT
Salinity Coefficient
0.0 (Default)
Setting range: 0 to 35 ppt
PRES
Atmospheric Pressure
Coefficient
760 (Default)
Setting range: 450 to 850
mmHg
CAL
Calibration Points
1 (Default)
1 point calibration
2
2 points calibration
UNIT
Measurement Unit
mg/L (Default)
Concentration unit
ppm
Concentration unit
%
% saturation
°C (Default)
Celsius
°F
Fahrenheit
°C
Temperature calibration
CAL
Temperature calibration
mode
HOLD
Auto-Hold
YES
Automatically freezes
stable readings
NO (Default)
Disable
OFF
Auto-off
YES (Default)
Automatically turns off
tester after 8 seconds of
idle
NO
Disable
RST
Reset
YES
Restores factory settings
NO (Default)
Disable
SETTING OF DEFAULT PARAMETERS
1. Press the CAL button to enter setup menu. The first display should be the current salinity coefficient.
2. Press CAL again to set the desired value. Press ENTER to confirm.
3. The tester should then display the atmospheric
pressure setting. The display should show ‘760mmHg’.