HOLD FUNCTION 
 

1. Pressing the HOLD key while in measurement mode will activate the hold function. The HOLD indicator will appear  
    on the display and further measurement readings will not be possible. 
2. Press the HOLD key again to disable the hold function. 
 
 

 
 

 

 
 
 
 
 
ELECTRODE SENSOR MAINTENANCE 

 
1. It is recommended to keep the sensitive membrane of the sensor wet. This is accomplished by periodically dipping  
    the sensor in distilled water or by using an electrode storage solution when the tester is not in use. 
2. If the instrument will be kept in storage for a long time, place the sensor in an electrode storage solution. 
3. Make sure to wash the sensor with distilled water before and after use. 
4. Do not reuse prepared buffer solutions.  
  

ELECTRODE REPLACEMENT 
 

One of the advantages of using the Kensington KS-EC-10L-E is that the sensor can be replaced instead of buying a 
new tester when the sensor fails during calibration. Removing and installing the sensor is easy and convenient and 
also increases the utility of the instrument as additional sensors can be purchased for different applications. To replace 
the sensor: 
 
1. Twist the electrode collar counter-clockwise and pull the electrode away from the instrument 
 
 
 
 
 
 
2. Align the new electrode into the same spot, push, and turn clockwise to keep it secure. 
 

 
 
 
 
 
 
 
 
 
 
 
 

Содержание KS-EC-10L-E

Страница 1: ... Remove it from the package 2 Remove the button cap found on the bottom of the tester as shown in the above image 3 Place the battery into its designated slot Take note of the proper position of the battery to ensure the polarities placed correctly 4 Replace the battery cap KEYPAD INDEX Key Function 1 Serves as the power button of the tester 2 Initiates Data Hold Press once to hold the current rea...

Страница 2: ...on 146 5 µS cm is required stir the electrode gently 3 Make sure there are no air bubbles trapped in the slot of the sensor 4 Press and hold the CAL key The tester will now enter into calibration mode 5 Press or to set the calibration value 6 Ensure the displayed value matches your calibration standard Press MEAS to confirm 7 The calibration value will then automatically flash three 3 times and th...

Страница 3: ... storage solution when the tester is not in use 2 If the instrument will be kept in storage for a long time place the sensor in an electrode storage solution 3 Make sure to wash the sensor with distilled water before and after use 4 Do not reuse prepared buffer solutions ELECTRODE REPLACEMENT One of the advantages of using the Kensington KS EC 10L E is that the sensor can be replaced instead of bu...

Страница 4: ... E Range 1 0 199 9µS cm Accuracy 0 1 FS Calibration points 1 point Calibration solutions 146 5µS cm Power off Manual or Automatic 8 minutes after last key pressed Temperature compensation 0 50 C Temperature coefficient 2 C Cell constant K 1 Normalization Temperature 25 C Hold function Manual Operating temperature 0 60 C 32 140 F Power requirements 3 1 5V G13A Batteries Dimensions 185 L 40 Dia mm W...

Отзывы: